It is recommended that you maintain a level of calibration appropriate to the datalogger
application. Consider the following factors when setting a calibration schedule:
l
The importance of the measurements
l
How long the datalogger will be used
l
The operating environment
l
How the datalogger will be handled
See also
You can download and print calibration certificates for many products you have purchased by
logging in to the Campbell Scientific website and going to:
https://www.campbellsci.com/calcerts
NOTE:
Note, you will need your product's serial number to access its certificate.
Watch an instructional
.
8.1.1 About background calibration
The datalogger uses an internal voltage reference to routinely self-calibrate and compensate for
changes caused by changing operating temperatures and aging. Background calibration
calibrates only the coefficients necessary to the running CRBasic program. These coefficients are
reported in the Status table as CalVolts(), CalGain(), CalOffset(), and CalCurrent().
Background calibration will be disabled automatically when the scan rate is too fast for the
background calibration measurements to occur in addition to the measurements in the program.
The
Calibrate()
instruction can be used to override or disable background calibration.
Disable background calibration when it interferes with execution of very fast programs and less
accuracy can be tolerated. With background calibration disabled, measurement accuracy over
the operational temperature range is specified as less accurate by a factor of 10. That is, over the
extended temperature range of –55 °C to 85 °C, the accuracy specification of ±0.08 % of reading
can degrade to ±0.8 % of reading with background calibration disabled. If the temperature of
the datalogger remains the same, there is little calibration drift when background calibration is
disabled.
8.2 Datalogger security
Datalogger security concerns include:
l
Collection of sensitive data
l
Operation of critical systems
8. Maintaining your datalogger
100
Содержание CR1000X
Страница 1: ...Revision 02 21 2019 Copyright 2000 2019 Campbell Scientific ...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 16: ...Appendix A Glossary 176 Appendix B Index 214 Table of Contents viii ...
Страница 20: ...l Smart sensors o SDI 12 o RS 232 o Modbus o DNP3 o TCP IP o RS 485 1 Data acquisition system components 4 ...
Страница 192: ......
Страница 237: ...wiring panel 5 Appendix B Index 220 ...