Appendix D. CNF4 Heater/Ventilator
D-8
Pull back wires for Pt-100 (grey, brown, green, and yellow), which are not in
use, and tie them around the TEMP cable using a cable tie or electrical tape to
avoid possible damage to the Pt-100, due to electrical short circuit.
Do not use the datalogger’s switched 12 V to simultaneously
power the heater and ventilator. Simultaneously powering
the heater and ventilator will exceed the current limit of the
switched 12 V channel. If the heater and ventilator need to
be used at the same time, connect the CNF4 to the 12V
channel instead of the switched 12 V channel and use an
external relay to switch the power on and off. Refer to
Section 4.1 of the CR1000 and CR3000 manual for details
on the 12V current source limits.
D.4 CR3000 Datalogger Program with
Heater/Ventilator Control
This example measures the four radiation outputs, thermistor temperature, and
controls the ventilator and heater using SW12V-1 and SW12V-2 channels on
the CR3000, respectively. In this example program, the ventilator and heater
can be turned on or off by manually setting the flag(1) and flag(2) high or low,
respectively. The program can be modified to include the conditional
statements to control the heater and ventilator based upon the environmental
parameters, such as light level and dew point temperature.
CRBasic Example D-1. CR3000 Datalogger Program with Heater/Ventilator Control
'CR3000 Series Datalogger
'
'CNR4 program
'This program measures CNR4 four-component net radiometer
'This program also measures the thermistor inside the CNR4
'In addition this program controls heater and ventilator
'using separate SW12V-1 and SW12V-2 channels
'The heater and ventilator are turned on/off by setting flag(1), and flag(2) high and low, respectively.
'
'
'User must enter the sensitivity values for all four probes in the program and save/compile
'prior to downloading it to the datalogger.
'Search for the text string "unique" to find places to enter the sensitivity values.
'
'Wiring Instructions
'
'ANALOG CHANNELS
'1H CNR4 Pyranometer Upper signal (red)
'1L CNR4 Pyranometer Upper signal reference (blue)
'gnd jumper to 1L
'
'2H CNR4 Pyranometer Lower signal (white)
'2L CNR4 Pyranometer Lower signal reference (thin black)
'gnd jumper to 2L
'
'3H CNR4 Pyrgeometer Upper signal (grey)
'3L CNR4 Pyrgeometer Upper signal reference (yellow)
'gnd jumper to 3L
'
'4H CNR4 Pyrgeometer Lower signal (brown)
'4L CNR4 Pyrgeometer Lower signal reference (green)