9.3 Recalibration
For quality assurance of the measured data, the manufacturer recommends the
CNR4 be recalibrated on a regular schedule by an authorized Kipp & Zonen
calibration facility.
The CNR4 should be recalibrated every two years. Alternatively, one can
check the sensor calibration by letting a higher standard run parallel to it over a
two-day period and, then, comparing the results. For comparison of
pyranometers, one should use a clear day. For comparison of pyrgeometers,
one should compare the nighttime results. If the deviations are greater than 6%,
the sensor should be recalibrated.
Please contact Campbell Scientific to obtain an RMA number for recalibration.