Zero Air Generator
13
9.
Invert the bottle assembly so the tubes can be more easily pushed in to the
molecular sieve.
10.
Tighten the two 1/8-in Swagelok
®
nuts.
11.
Tighten the four screws that clamp the lid to the bracket.
12.
Reattach the box lid and tighten the screws until snug.
7.3 Replacing Filter
The Zero Air Generator’s inlet filter is likely to last several years with normal
usage. Heavy use in dirty environments may eventually clog the filter causing
a reduction of air flow. If the inlet filter becomes plugged it may be replaced
with pn 29998 through the following steps.
1.
Loosen the three screws and remove the box lid.
2.
Twist the fittings on each side of the filter to remove the filter.
3.
Replace the filter with pn 29998.
4.
Tighten the fittings securely to avoid leakage.
5.
Reattach the box lid and tighten the screws until snug.