![Campbell 31022 Скачать руководство пользователя страница 11](http://html.mh-extra.com/html/campbell/31022/31022_instruction-manual_489145011.webp)
Zero Air Generator
3
4.1 Accessories
4.1.1 Other Accessories
Plastic tubing
Bev-A-Line IV plastic tubing with an outer diameter of 1/4 in and inner
diameter of 1/8 in, is available as pn 7399. The tubing remains flexible even at
cold temperatures and holds up well in applications that require repeated
handling and flexing. Campbell Scientific recommends this tubing for the
plumbing connection between the Zero Air Generator and the instrument under
test.
4.1.2 Replacement Parts
AA alkaline cells
The power that operates the small pump of the Zero Air Generator is supplied
by a pair of AA alkaline cells. Campbell Scientific offers pn 26064 which is a
long-life 1.5 V, AA alkaline battery.
Molecular sieve
The Zero Air Generator scrubs an air sample of CO
2
and water vapor with a
13X molecular sieve. The sieve requires periodic replacement. Campbell
Scientific offers pn 27450, which is a bottle containing 250 g of 13X molecular
sieve beads ranging from 1.6 to 2.5 mm diameter.
Two 250 g bottles are required to replace the sieve.
Filter
The original filter of the Zero Air Generator can be replaced by the filter
available as pn 29998. The filter is an in-line, 2.5 cm (1.0 in) diameter, PTFE
membrane filter of 3-micron pore size with Luer lock connections.
Swagelok
®
plug
Spare 1/4-in Swagelok
®
plugs are available as pn 15891. This part is used to
plug a fitting when the accompanying tube is disconnected. Plugging the
fittings is necessary to keep the fittings clean and to avoid consumption of the
molecular sieve during storage. It is strongly recommended to replace plugs in
the event that those provided with the Zero Air Generator are lost or become
damaged.
NOTE