110PV Surface Temperature Probe
6
Temperature Range:
–40 to 135 °C
Survival Range:
–50 to 140 °C
Accuracy
1
Worst Case:
±0.2
°
C (–40 to 70 °C)
±0.5
°
C (71 to 105 °C)
±1
°
C (106 to 135 °C)
Maximum Steinhart-Hart
Linearization Error:
0.0024 °C at –40 °C
Time Constant (average)
Still Air:
252 s
Surface:
25 s
Water Submersion Depth:
50 ft (21 psi)
Maximum Cable Length:
304.8 m (1000 ft)
Disk Diameter:
2.54 cm (1.0 in)
Overall Probe Length:
6.35 cm (2.5 in)
Overmolded Joint Dimensions
Width:
1.12 cm (0.44 in)
Height:
1.47 cm (0.58 in)
Length:
5.72 cm (2.25 in)
Cable Diameter:
0.622 cm (0.245 in)
Material
Disk:
Anodized Aluminum
Cable Jacket:
Santoprene
®
Cable/Probe Connection:
Santoprene
®
Weight:
90.7 g (0.2 lb) with 3.2 m (10.5 ft) cable
1
shows the Steinhart-Hart error. FIGURE
temperature uncertainty. The overall probe accuracy is a combination of the
thermistor’s interchangeability specification and the accuracy of the bridge resistor.
The major error component is the interchangeability specification (tolerance) of the
thermistor. The bridge resistor has a 0.1% tolerance with a 10 ppm temperature
coefficient. Effects of cable resistance is discussed in Section
(p. 14)
.
Содержание 110PV
Страница 2: ......
Страница 6: ......
Страница 26: ...110PV Surface Temperature Probe 18 ...
Страница 28: ......
Страница 34: ...Appendix C Probe Material Properties C 2 ...
Страница 35: ...Appendix C Probe Material Properties C 3 ...
Страница 36: ...Appendix C Probe Material Properties C 4 ...
Страница 37: ......