110PV Surface Temperature Probe
3
4.
Select
Datalogger Model
and
Scan Interval
(default of
5
seconds is OK
for most applications). Click
Next
.
5.
Under the
Available Sensors and Devices
list, select the
Sensors
|
Temperature | Soil Moisture | 110PV
. Click
to move the selection
to the selected device window. Enter the
Cable Resistance
. This value is
unique for each 110PV, and is printed on the heat shrink label attached to
the sensor cable. The surface temperature defaults to degree C. This can
be changed by clicking the
Temperature
box and selecting one of the
other options.