2.3.4 Printing results
If a printer is installed and switched on, it is possible to print out the test results (without
saving before).
Press F3 key.
The complete data set is printed: date, time, Code-No. and
test result. Printing example:
2006-07-01 14:53:09
Test No.:1
Code-No.:1
2,13 NTU*
The Test No. is a system number for data storage capacity
and is set automatically if a test result is stored. It appears
only at the print out.
The data set is stored with date, time, Code-No. and test
result.
The display shows:
The test result is then shown again.
Note:
The display shows the number of free data sets.
If there are less than 30 data sets free the display shows:
Clear the memory as soon as possible (see chapter 2.4.4
Delete stored results). If memory capacity is used up it
would be impossible to save additional test results.
Stored!
F3
Storage:
only 29 free records left
Storage: 900
free records left