1.4.2 Cleaning of vials & Sampling vessels
Vials, caps and sampling vessels should be cleaned thoroughly after each test to avoid influ-
ences. Minor residuals can cause errors.
Residuals:
According to the type of measured sample it is necessary to clean the vial with different
cleaning operations.
• Replace scratched vials immediately.
• Rinse vials continuous with deionized water after each measurement.
• Clean all glassware continuous with laboratory detergent and rinse with deionized water
thoroughly.
• Clean heavy conterminations by filling the vials with 1:1 HCL followed by multiple rinses with
distilled or deionized water.
• Allow vials to air dry.
• Touch vials at the top only to minimize dirt and fingerprints.
• Wipe the vials with a lint-free cloth to remove waterdrops and fingerprints.