· Turn off the meter and remove the test leads from the meter 

before opening the back cover or the case.
· Do not operate the meter with the back cover or portions of 

the case removed or loosened.
· To avoid false readings, which could lead to possible electric 

shock or personal injury, replace the battery as soon as the 

low battery indicator (         )  appears.
· To avoid electric shock, do not contact any open conductor 

with hands or skin. Do not ground yourself while using the 

meter.
· Remaining endangerment:

  When an input terminal is connected to dangerous live 

potential, it is to be noted that this potential can occur at all 

other terminals.
· 

CAT III 

- Measurement category 

III 

is for measurements 

performed in the building installation. Examples are 

measurements on distribution boards, circuit breakers, 

wiring, including cables, bus-bars, junction boxes, switches, 

socket-outlets in the fixed installation, and equipment, for 

example, stationary motors with permanent connection to the 

fixed installation. Do not use the meter for measurements 

within Measurement Categories IV.

CAUTION

To avoid possible damage to the meter or to the equipment 

being tested, follow these guidelines:
· Disconnect circuit power and discharge all capacitors before 

measuring resistance, diode, temperature, or continuity.
· Use the proper terminals, function, and range for measure-

ments.
· Before measuring current, check the meter’s fuses and turn 

off the power to the circuit before connecting the meter to the 

circuit.

Resolution

Accuracy

200Ω

2kΩ

20kΩ

200kΩ

2MΩ

± (1.2% + 5)

Range

0.1Ω

0.001kΩ

0.01kΩ

0.1kΩ

0.001MΩ

Note:  

1. Use K type thermocouple

2. Accuracy does not include error of the thermocouple probe.

3. Accuracy specification assumes ambient temperature 

is stable to ±1˚C. For ambient temperature changes of 

±5˚C, rated accuracy applies after 1 hour.

[1] 

If the temperature being measured is out of the range of 0ºC to 

1000ºC or 0ºF to 1400ºF, the display may show a reading; but the 

measurement error may be large or the thermocouple may be damaged.

Frequency

Output Level

(squarewave)
(sinewave)

about 3.0Vp-p
about 0.8Vp-p

Range

about 50Hz
about 1kHz

Signal Output (EM830D only)

The working voltage of the 

battery will be shown on the 

display so that the quality of 

the battery can be judged

Description

Remark

1.5V

9V

Test Current:

about 20mA

Test Current:

about 5mA

Range

Battery (EM830 only)

Overrange

Indication

"OL" shown

on the display

Resolution

Accuracy

0˚C - 1000˚C

0˚F - 1400˚F

± (2.0% + 3)
± (2.0% + 5)

Range

1˚C

1˚F

Temperature (EM830C only)

Overrange

Indication

[1]

Содержание EM830 Series

Страница 1: ...manship for a period of one year Any instrument found defective within one year from the delivery date and returned to the factory with transpor tation charges prepaid will be repaired adjusted or replaced at no charge to the original purchaser This warranty does not cover expandable items such as battery or fuse If the defect has been caused by a misuse or abnormal operation conditions the repair...

Страница 2: ...the rated voltage as marked on the meter between terminals or between any terminal and earth ground Before use verify the meter s operation by measuring a known voltage When measuring current turn off circuit power before connecting the meter in the circuit Remember to place the meter in series with the circuit When servicing the meter use only specified replace ment parts Use caution when working...

Страница 3: ...T III Measurement category III is for measurements performed in the building installation Examples are measurements on distribution boards circuit breakers wiring including cables bus bars junction boxes switches socket outlets in the fixed installation and equipment for example stationary motors with permanent connection to the fixed installation Do not use the meter for measurements within Measu...

Страница 4: ...ouble insulation or reinforced insulation INTRODUCTION The EM830 series meters are compact 3 digit digital multimeters designed to measure DC and AC voltage DC current resistance and diode Some models also have continuity temperature or battery measurement function or signal output function These meters feature low batery indication and full range overload protection They are ideal measurement too...

Страница 5: ...tion when the meter is not in use 3 COM Terminal Plug in the black test lead connector for all measure ments except temperature measurements For tempera ture measurements this COM terminal is a plug in connector for the negative plug of the K type thermocouple 4 VmAΩ Terminal Plug in the red test lead connector for all measurements except temperature measurements and current measurements 200mA For...

Страница 6: ...0 C Relative Humidity 75 RH Storage Environment Temperature 10 C to 50 C Relative Humidity 85 RH Size 126mm x 70mm x 26mm Weight about 135g including battery SPECIFICATIONS Accuracy is specified for a period of one year after calibration at 18 C to 28 C with relative humidity up to 75 Accuracy specifications take the form of of Reading Number of Least Significant Digits DC Voltage Input Impedance ...

Страница 7: ...e 0 01μA 0 1μA 1μA 0 01mA 0 1mA 0 01A Overload Protection 250mA 300V Fast fuse for protection for the inputs of terminal 10A 300V Fast fuse for protection for the inputs of 10A terminal Max Allowable Input Current 10A For measurements 2A measurement duration 10 secs and interval 15 minutes Note Only the EM830B has 20μA and 200μA ranges VmAΩ Max Open Circuit Voltage about 1V Resistance Overrange In...

Страница 8: ...o 1400ºF the display may show a reading but the measurement error may be large or the thermocouple may be damaged Frequency Output Level squarewave sinewave about 3 0Vp p about 0 8Vp p Range about 50Hz about 1kHz Signal Output EM830D only The working voltage of the battery will be shown on the display so that the quality of the battery can be judged Description Remark 1 5V 9V Test Current about 20...

Страница 9: ...sured 4 Read the display reading The polarity of the red test lead connection will be indicated as well Note To avoid electric shock or damages to the meter do not apply a voltage higher than 300V between terminals The display shows the approx forward voltage drop of the diode Open Circuit Voltage about 2 2V Test Current about 1mA Description Range Diode Test The built in buzzer will sound if the ...

Страница 10: ...rent 1 Connect the black test lead to the COM terminal Connect the red test lead to the terminal if the current to be measured is less than 200mA If the current is between 200mA and 10A connect the red test lead to the 10A terminal instead 2 If the red test lead is connected to the terminal set the range switch to the desired range position in the area If the red test lead is connected to the 10A ...

Страница 11: ...surements 2 When the input is not connected i e at open circuit OL will be displayed as an overrange indication 3 Before making in circuit resistance measurement disconnect all power to the circuit to be tested and discharge all capacitors thoroughly Diode Test 1 Connect the black test lead to the COM terminal Connect the red test lead to the terminal Note The polarity of the red lead is positive ...

Страница 12: ...e K type thermo couple to the COM terminal and the plug of the K type thermocouple to the terminal 2 Set the range switch to ºC position for celsius temperature measurements or to the ºF position for fahrenheit temperature measurements 3 Connect the sensing end of the thermocouple to the object to be measured 4 Wait a moment until thermal equilibrium between the object and the thermocouple probe i...

Страница 13: ...to repair or service the meter Store the meter in a dry place when not in use Don t store it in an intense electromagnetic field environment General Maintenance Periodically wipe the case with a damp cloth and a little mild detergent Do not use abrasives or solvents Dirt or moisture in the terminals can affect readings Clean the terminals as follows 1 Set the range switch to OFF position and remov...

Страница 14: ...e the exhausted battery with a new one of the same type 9V battery 6F22 or equivalent Reinstall the back cover and the screws To replace fuse remove the screws on the back cover and remove the back cover replace the blown fuse with a new one of the same ratings Reinstall the back cover and its screws This meter uses two fuses F 1 250mA 300V Fast fuse Min Interrupt Rating 1500A Ø5X20mm F 2 10A 300V...

Страница 15: ... período de un año Todo instrumento que se encontrara con defectos dentro de un año desde la fecha de entrega y se retornara a la fábrica con cargos de transporte prepagos se reparará ajustará o reemplazará sin caragos para el comprador original Esta Garantía no cubre ítems expansivos tales como batería o fusible Si el defecto fuera causado por mal uso o condiciones de funcionamiento anormales la ...

Страница 16: ...ierra Antes de usarlo verifique el funcionamiento del medidor mediante la medición de un voltaje conocido Cuando mida la corriente desconecte la potencia del circuito antes de conectar el medidor en el circuito Recuerde colocar el medidor en serie con el circuito Cuando haga reparaciones en el medidor use solamente piezas de reemplazo específicas Tenga cuidado cuando trabaje con un voltaje por enc...

Страница 17: ...del edificio Por ejemplo mediciones en tableros de distribución interruptores de circuito cableados incluidos cables barras de conducción cajas de unión interruptores tomacorrientes y equipamiento por ejemplo motores fijos con conexión permanente a la instalación fija No use el medidor para mediciones dentro de las Categorías IV de medición PRECAUCIÓN Para evitar un posible daño al medidor o al eq...

Страница 18: ...mpactos digitales de 3 dígitos diseñados para medir voltaje de CC y CA corriente CC resistencia y diodo Algunos modelos también tienen función de medición de continuidad temperatura o batería o función de generación de señal Estos medidores presentan indicación de bacteria baja y protección de sobrecarga de rango completo Son herramientas de medición ideales Excepto donde se indique las descripcio...

Страница 19: ... mediciones de temperatura Para las mediciones de temperatura esta terminal COM es un conector que se enchufa para la toma negativa de la termocupla tipo K 4 Terminal Enchufe el conector de guía de prueba rojo para todas las mediciones con excepción de las mediciones de temperatura y de corriente 200mA Para mediciones de temperatura es un conector que se enchufa para la toma positiva de la termocu...

Страница 20: ...C Humedad relativa 75 RH Entorno de almacenamiento Temperatura 10 C a 50 C Humedad relativa 85 RH Tamaño 126mm x 70mm x 26mm Peso alrededor de 135g batería incluida ESPECIFICACIONES La precisión está especificada para un periodo de un año luego de la calibración a 18 C a 28 C con una humedad relativa de hasta 75 Especificaciones sobre precisión toman la forma de de lectura Cant de dígitos menos si...

Страница 21: ...de 1V Resistance Impedancia de entrada 10MΩ Rango de frecuencia 40Hz a 400Hz Respuesta Promedio calibrada en rms de onda sinusoidal Voltaje máximo de entrada permitido 300V Resolución Precisión Indicación de sobrerango 200V 300V 1 2 10 Rango 0 1V 1V En el visualizador se muestra OL Resolución Precisión Indicación de sobrerango 20μA 200μA 2000μA 20mA 200mA 10A 1 2 5 2 0 5 Rango 0 01μA 0 1μA 1μA 0 0...

Страница 22: ...e de trabajo de la batería se mostrará en el visualizador como para que se pueda juzgar la calidad de la batería Descripción Comentario 1 5V 9V Corriente de prueba aprox e 20mA Corriente de prueba aprox 5mA Rango Batería EM830 solamente Generación de señal EM830D solamente onda cuadrada onda sinusoidal aprox 3 0Vp p aprox 0 8Vp p aprox 50Hz aprox 1kHz Frecuencia Niveldegeneración Rango Temperatura...

Страница 23: ... la guía de prueba roja Nota Para evitar choque eléctrico o daños al medidor no aplique un voltaje mayor a 300V entre terminales Medición de voltaje CA 1 Conecte la guía de prueba negra a la terminal COM y la guía de prueba roja a la terminal 2 Fije el interruptor de rango al rango deseado Si la magnitud de voltaje a ser medido no se conoce de antemano fije el interruptor de rango al rango más alt...

Страница 24: ...fije el interruptor de rango en la posición de rango 3 Desconecte la energía al circuito a ser probado Luego descargue todos los capacitores 4 Interrumpa el trayecto del circuito a ser medido luego conecte las guías de prueba en serie con el circuito 5 Conecte la energía al circuito luego lea el visualizador También se indicará la polaridad de la conexión de la guía de prueba roja Nota 1 Si la mag...

Страница 25: ...a negra a la terminal COM Conecte la guía de prueba roja a la terminal Nota La polaridad de la guía red es positiva 2 Fije el interruptor de rango en la posición deseada 3 Conecte la guía de prueba roja al ánodo del diodo a ser probado y la guía de prueba negra al cátodo del diodo 4 El visualizador muestra la caída del voltaje hacia adelante aproximado del diodo Si la conexión se invierte se mostr...

Страница 26: ... para las mediciones de temperatura celsius o en la posición ºF para la mediciones de temperatura Fahrenheit 3 Conecte el extremo de detección de la termocupla al objeto a ser medido 4 Espere un momento hasta que se llegue al equilibrio termal entre el objeto y la sonda de la termocupla luego lea la lectura del visualizador Prueba de batería EM830 solamente 1 Conecte la guía de prueba negra a la t...

Страница 27: ...o técnico al medidor Guarde el medidor en un lugar seco cuando no se use No lo guarde en un ambiente de campo electromagnético intenso Mantenimiento General Limpie periódicamente la caja con un trapo húmedo y detergente suave No use abrasivos ni solventes La suciedad o la humedad en las terminales pueden afectar las lecturas Limpie las terminales de la siguiente manera 1 Fije el interruptor de ran...

Страница 28: ... con una nueva del mismo tipo batería de 9V 6F22 o equivalente Reinstale la cubierta posterior y los tornillos Para reemplazar fusibles quite los tornillos de la cubierta posterior y quita la cubierta posterior reemplace el fusible dañado con un nuevo con las mismas especificaciones Reinstale la cubierta posterior y sus tornillos Este medidor usa dos fusibles F 1 Fusible rápido de 250mA 300V Espec...

Отзывы: