Linomat 5
Instruction manual, March 2016
Page 16
Set the LCD contrast
Start the user dialog according to the procedure above. The dis-
play now shows the current baud rate.
Press the
ENTER
key twice to move on to the LCD contrast se-
lection display.
The current contrast setting is displayed.
Press the
or
key to change the setting.
Press the
ENTER
key to accept the current selection.
Press the
RESET
key to restart the instrument.
5.3
Handling of the HAMILTON-syringe
To achieve a maximum lifetime for the gas-tight HAMILTON-
syringe the following points must be observed:
•
Ensure extremely clean conditions when handling the syringe
body and plunger.
•
Never lubricate the plunger.
•
The Teflon tip of the plunger is very sensitive. Even slightest
scratches can cause a leak.
•
Plungers and syringe bodies are exchangeable.
•
To fill the sample dosage syringe make sure the needle is not
pushed against the bottom of vial and that no air bubbles re-
main in the needle.
The plunger must not be exposed to boiling water.
The following replacement plunger can be obtained:
022.7830 Replacement plunger for 100
µ
L syringe
022.7831 Replacement plunger for 500
µ
L syringe
5.4
Adjusting the spray gas
Spray gas flow is controlled by a gas pressure regulator, factory
adjusted to 1.0 bar. It may be suitable to increase this pressure to
achieve better spray quality when using highly viscous liquids.
Procedure:
Turn the spray pressure regulator (1) clockwise to increase
pressure. The pressure can be read from an external gauge you
can attach at (2). Adjust pressure to 1.5 bar.