6.2 Front Panel Calibration Screens
The calibration screens for output or measurement calibration can be selected from the MENU 3
light
ems with three-phase output capability, use the PHASE key on the front panel to
ame
icated above.
screen. (Press MENU button several times to toggle to MENU 3 screen.)
To select the OUTPUT CALIBRATION screen press the
↑
or
↓
key several times to high
OUTPUT CAL. Then press the ENTER key. This will bring up the PASSWORD screen. To
prevent unauthorized access to calibration data, a password must be entered to access any
calibration screen. The calibration password is always "
5000
" and may be entered using the
numeric keypad. Once entered, the calibration screens remain accessible until the MX unit is
powered down.
Type 5000 and press the ENTER key to show the OUTPUT CALIBRATION screen.
On MX syst
select the phase to be calibrated.
To select the MEASUREMENT CALIBRATION screen, follow the same steps as outlined above
but select the MEASUREMENT CAL entry instead of OUTPUT CAL. If another CALIBRATION
screen has been accessed since power-up, no password is needed. Otherwise, enter the s
password as ind
MX Series
157
Содержание MX30-1
Страница 4: ...User Manual Rev AJ California Instruments 4 MX Series ...
Страница 33: ...User Manual Rev AJ California Instruments MX Series 33 ilter option schematic Figure 2 12 MX output f ...
Страница 34: ...User Manual Rev AJ California Instruments Figure 2 13 MX45 Output Noise 10 KHz 1 MHz 34 MX Series ...
Страница 50: ...User Manual Rev AJ California Instruments Figure 3 11 MX135 or MX135 MB Output Wiring Rear view 50 MX Series ...
Страница 147: ...User Manual Rev AJ California Instruments Output Status Available on top assembly 7003 427 only MX Series 147 ...
Страница 148: ...User Manual Rev AJ 148 California Instruments MX Series Figure 5 2 MX Series Detailed Block Diagram ...
Страница 149: ...User Manual Rev AJ MX Series California Instruments 149 Figure 5 3 Power Module Detailed Block Diagram ...
Страница 195: ...User Manual Rev AJ California Instruments Figure 9 4 Voltage Modulation MX Series 195 ...