User Manual – Rev B
FCS Series II
3.7 Basic Initial Functional Test
CAUTION:
Work carefully when performing these tests; hazardous voltages
are present on the input and output during this test.
Figure 3-7: Functional Test Setup
Refer to Figure 3-7 for the required functional test set up. Proceed as follows to perform a basic
functional check of the power system:
1. Verify the correct AC line input rating on the nameplate of the FCS unit(s) and make sure
ing
n Table
with –
has sufficient power dissipation capability for full
the correct three-phase line voltage is wired to the input of the FCS cabinet before apply
input power.
2. Connect a suitable resistive or other type load to the output of the power source. Suggested
load values for the low voltage range rounded up to the nearest 0.1 Ohm are shown i
3-13.
For the high voltage range, the resistor values must be multiplied by four. For models
HV or –EHV voltage range option, recalculate these values by adjusting for the voltage
range value. Make sure the power resistor
California Instruments
42
Содержание FCS Series II
Страница 4: ...User Manual Rev B FCS Series II California Instruments 4...
Страница 27: ...User Manual Rev B FCS Series II California Instruments 27...
Страница 28: ...User Manual Rev B FCS Series II Figure 3 1 Rear Panel Connector Locations California Instruments 28...
Страница 35: ...User Manual Rev B FCS Series II Figure 3 4 FCS54 3 3 Wiring Diagram Three Phase mode California Instruments 35...
Страница 36: ...User Manual Rev B FCS Series II Figure 3 5 FCS54 3 1 Wiring Diagram Single Phase mode California Instruments 36...
Страница 89: ...FCS Series II 89 Figure 5 1 FCS System Block Diagram User Manual Rev B California Instruments...
Страница 92: ...User Manual Rev B FCS Series II Figure 5 2 Power Amplifier Exploded View California Instruments 92...
Страница 95: ...User Manual Rev B FCS Series II Figure 6 1 Calibration Setup California Instruments 95...
Страница 109: ...User Manual Rev B FCS Series II Figure 7 1 Front panel location Figure 7 2 Fuse Locations California Instruments 109...
Страница 118: ...User Manual Rev B FCS Series II ure 8 4 Voltage Modulation Frequency characteristics Fig California Instruments 118...