Setting
Description
Alarm
Output
(continued)
Alarm Trigger
HI:
The alarm is triggered if the temperature
is higher than the Set Point.
LO:
The alarm is triggered if the temperature
is lower than the Set Point.
Output Mode
Active HI:
In an alarm condition, the alarm wire AL
will sink current to ground through the
attached load (e.g. a relay).
Active LO:
In a non-alarm condition, the alarm wire
AL will sink current as above.
Temperature
Reference
Measured:
The temperature of the target object.
Internal:
The temperature inside the housing of the
sensor. This can be used to indicate the
ambient temperature where the sensor is
mounted.
Display
Displayed
Temperature
RAW:
Measured temperature without averaging
or hold processing
FIL:
Includes averaging and hold processing
AVG:
Includes averaging only
This setting does not affect the sensor’s outputs.
Temperature
Units
Select Celsius or Fahrenheit. Units are changed
throughout the sensor’s interface.
Screen Saver
If enabled, the sensor’s display turns itself off after 1
minute of inactivity. Press any key to turn the display on
again. The display will still turn on and flash if the alarm
is triggered.
Rotation
Change this setting from “Normal” to “Inverted” to make
the display easy to read if the sensor is installed upside-
down.
Reset
Alarm
In Manual Reset Mode, after the alarm has been triggered and
acknowledged, this function allows the alarm to be triggered again.
Enter PIN /
Lock
Sensor
Enter the PIN to unlock the sensor and change the settings. The PIN is
1234. When configuration is complete, lock the settings again by selecting
Lock Sensor.