IMPORTANT

Be aware of the following when using the sensor:

If the sensor is exposed to significant changes in ambient temperature (hot to cold, or cold
to hot), allow 20 minutes for the temperature to stabilise before taking or recording
measurements.

Do not operate the sensor near large electromagnetic fields (e.g. around arc welders or
induction heaters). Electromagnetic interference can cause measurement errors.

Wires must be connected only to the appropriate terminals.

Ensure the sensor is grounded at one point: either the cable shield termination, or
the mounting screws, but not both.

When the power is switched off, allow at least 15 seconds before switching the power back
on.

LENS CLEANING

Keep the lens clean at all times. Any foreign matter on the lens would affect measurement
accuracy. Blow off loose particles (if not using the air purge accessory) with an air ‘puffer’.

MAINTENANCE

Our customer service representatives are available for application assistance, calibration,
repair, and solutions to specific problems. Contact our Service Department before returning
any equipment. In many cases, problems can be solved over the telephone. If the sensor is not
performing as it should, try to match the symptom below to the problem. If the table does not
help, call Calex for further advice.

TROUBLESHOOTING

Symptom

Probable Cause

Solution

No output

No power to sensor

Check power supply and wiring

Inaccurate measured
temperature

Target too small for
sensor’s field of view

Ensure the sensor’s view is completely filled
by the target. Position the sensor closer to
the target to measure a smaller area.

Target is a reflective
metal surface

Measure a non-reflective area, or paint or
coat a measurable area of the target to
make it non-reflective

Field of view
obstruction

Remove obstruction; ensure sensor has a
clear view of target

Dust or condensation
on lens

Ensure lens is clean and dry. Clean gently
with a soft lens cloth and water. If problem
recurs, consider using an air purge collar.

Incorrect wire
connections

Check wire colour codes

Erroneous
temperature (voltage
output)

Output temperature
scale mismatch

Re-scale input temperature range on
measurement instrument to match sensor

Erroneous
temperature
(thermocouple
output)

No Cold Junction
Compensation (CJC) or
wrong type of
extension cable

Enable CJC on measurement instrument;
ensure extension cable and connectors are
of the correct thermocouple type

7

Содержание PyroNFC Series

Страница 1: ...PyroNFC Series Operator s Guide...

Страница 2: ...Adjustable via app Emissivity Setting Range 0 2 to 1 0 Max Temperature Span Linear Output 1000 C Min Temperature Span Linear Output 100 C Spectral Range 8 14 m Max Supply Voltage 28 V DC Min Supply V...

Страница 3: ...perate in ambient temperatures from 0 C to 80 C Avoid thermal shock Allow 20 minutes for the unit to adjust to large changes in ambient temperature ATMOSPHERIC QUALITY Smoke fumes or dust can contamin...

Страница 4: ...before switching it back on MECHANICAL INSTALLATION All sensors come with a 1 metre cable and two mounting screws The sensor can be mounted on brackets or cut outs of your own design or you can use t...

Страница 5: ...urs and Tags 14 0 13 5 31 0 2 x Hole 3 2 Section A A A A 21 5 THIS FACE AGAINST SENSOR AIR OUT Air Inlet Hose barb 5 mm Touch NFC smartphone here Field of view see OPTICS PWR PWR OP 0 V 21 5 31 0 29 0...

Страница 6: ...ntinuously while the phone is touching the sensor On iOS the temperature is updated once each time the user taps update in the app The sensor must be powered to read the measured temperature with the...

Страница 7: ...can be solved over the telephone If the sensor is not performing as it should try to match the symptom below to the problem If the table does not help call Calex for further advice TROUBLESHOOTING Sym...

Страница 8: ...K thermocouple fixed Temperature Low 0 C to 900 C 0 C 0 C fixed Temperature High 100 C to 1000 C 1000 C 1000 C fixed Alarm Trigger High Low Off High High Alarm Monitor Temperature Target Internal Targ...

Отзывы: