Calex PyroNFC Series Скачать руководство пользователя страница 2

The PyroNFC is a non-contact infrared temperature sensor with Near Field Communication
(NFC) capability. It has linear voltage and open collector alarm outputs for connection to
industrial process instrumentation.

By touching an NFC-enabled smartphone against the back of the sensor, the user can
configure the sensor’s settings, and read the measured temperature, via a free app.

SPECIFICATIONS

General

Temperature Range

0 to 1000°C

Outputs

Model PN151: 0-5 V DC, 1-5 V DC or 0-10 V
DC output, linear with measured temperature,
configurable and rescalable via NFC

Model PN151K: Type K Thermocouple

Field of View

15:1 (see OPTICS)

Accuracy

± 1.5% of reading or ± 1.5°C, whichever is greater

Repeatability

± 0.5% of reading or ± 0.5°C, whichever is greater

Response Time

125 ms

Configuration

Via iOS or Android app using NFC-equipped
device (e.g. smartphone or tablet)

Emissivity Setting

Adjustable via app

Emissivity Setting Range

0.2 to 1.0

Max Temperature Span (Linear Output)

1000°C

Min Temperature Span (Linear Output)

100°C

Spectral Range

8-14 µm

Max. Supply Voltage

28 V DC

Min. Supply Voltage (at Sensor)

12 V DC (for 10 V output)
6 V DC (for 5 V output)

Max Current Draw

7 mA

Warm Up Time (PN151K)

Allow 10 seconds after power-on before a reading
is observed.

Environmental & Conformity

Environmental Rating

IP65

Ambient Temperature Range

0°C to 80°C

Relative Humidity

95% max. non-condensing

Electromagnetic Compatibility (EMC)

EN61326-1, EN61326-2-3 (Electrical Equipment
for Measurement, Control and Laboratory Use -
EMC Requirements - Industrial)

RoHS Compliant

Yes

App

Configurable Parameters

Temperature range
Linear voltage output type and scale
Alarm output threshold and hysteresis
Emissivity setting
Reflected temperature

Temperature Units

°C / °F

Signal Processing

Averaging Period (0.125 to 60 seconds)
Peak / Valley Hold
Hold Period (0.125 to 120 seconds)

Real Time Temperature Reading

Hold NFC device against sensor for continuous in- app
temperature updates

2

Содержание PyroNFC Series

Страница 1: ...PyroNFC Series Operator s Guide...

Страница 2: ...Adjustable via app Emissivity Setting Range 0 2 to 1 0 Max Temperature Span Linear Output 1000 C Min Temperature Span Linear Output 100 C Spectral Range 8 14 m Max Supply Voltage 28 V DC Min Supply V...

Страница 3: ...perate in ambient temperatures from 0 C to 80 C Avoid thermal shock Allow 20 minutes for the unit to adjust to large changes in ambient temperature ATMOSPHERIC QUALITY Smoke fumes or dust can contamin...

Страница 4: ...before switching it back on MECHANICAL INSTALLATION All sensors come with a 1 metre cable and two mounting screws The sensor can be mounted on brackets or cut outs of your own design or you can use t...

Страница 5: ...urs and Tags 14 0 13 5 31 0 2 x Hole 3 2 Section A A A A 21 5 THIS FACE AGAINST SENSOR AIR OUT Air Inlet Hose barb 5 mm Touch NFC smartphone here Field of view see OPTICS PWR PWR OP 0 V 21 5 31 0 29 0...

Страница 6: ...ntinuously while the phone is touching the sensor On iOS the temperature is updated once each time the user taps update in the app The sensor must be powered to read the measured temperature with the...

Страница 7: ...can be solved over the telephone If the sensor is not performing as it should try to match the symptom below to the problem If the table does not help call Calex for further advice TROUBLESHOOTING Sym...

Страница 8: ...K thermocouple fixed Temperature Low 0 C to 900 C 0 C 0 C fixed Temperature High 100 C to 1000 C 1000 C 1000 C fixed Alarm Trigger High Low Off High High Alarm Monitor Temperature Target Internal Targ...

Отзывы: