INSTALLATION
The installation process consists of the following stages:
Preparation
Mechanical installation
Electrical installation
Please read the following sections thoroughly before proceeding with the installation.
PREPARATION
Ensure that the sensor is positioned so that it is focused on the target only.
ACCESSORIES
A range of accessories to suit different applications and industrial environments is available.
These may be ordered at any time and added on-site. The accessories consist of the
following parts.
Fixed and adjustable mounting brackets
Air purge collar
Laser sighting tool
OPTIONS
The following options are available. Options are factory installed and may only be ordered at
the same time as the PyroEpsilon sensor.
Air/water cooled housing
Certificate of calibration
Longer cable (30 m max.)
OPTICAL CHART
The optical chart below indicates the nominal target spot diameter at any given distance from
the sensing head and assumes 90% energy.
1000
103
12
18
M16 x 1mm
D:S 15:1
D:S 2:1
D:S 30:1
11.9
45.2
78.6
0.5
1.8
3.1
0
19.7
39.4
0
500
1000
11.9
0.5
61.9
111.9
2.4
4.4
4
8
0
0
5.0
0.20
100
3.9
12.5
0.49
200
7.9
100
200
11.9
28.6
45.2
0.5
1.1
1.8
0
19.7
39.4
0
500
1000
11.9
0
0.5
0
(inches)
Measured Spot Diameter
(mm)
Distance: Sensor to object (inches)
Distance: Sensor to object (mm)
Sensor
CORRECT
INCORRECT
Background
Target greater
than spot size
Target smaller
than spot size
PE21
PE151
PE301
PECF
english
www.digiparts.ch