CT4432 10 kV High-Voltage Oscilloscope Probe User Manual
Specifications
The probe and oscilloscope should be warmed up for at least 20 minutes before any testing and the
environmental conditions should not exceed the probe’s specified limits.
All entries included in the following tables are characteristics unless otherwise stated.
Toutes les entrées incluses dans les tableaux suivants sont des caractéristiques, sauf
indication contraire.
Table 2
Safety Specifications
Parameter
Condition
IEC/EN 61010-031:2015
Measurement Category
other than
II, III, or IV (formally CAT I)
Table 3
Electrical Specifications (Not Warranted)
Parameter
Characteristic
Maximum Input Voltage
10 kV (DC + ACpeak)
7 kV (ACrms)
Maximum Loading Current
100 µA
Attenuation
1000x
Bandwidth
DC to 40 MHz (-3 dB)
Rise Time
≤ 8.8 ns
Input Resistance
100 MΩ ± 5%
Input Capacitance
2.0 pF
Accuracy
± 3% (below 10 kV) DC
± 3% at 1 kHz VAC
± 3 dB 0 to 40 MHz VAC
Temperature Coefficient
≤ 200 ppm/°C
Compensation Range
5 pF to 50 pF
NOTE
REMARQUE