Cal Test Electronics CT3684 Скачать руководство пользователя страница 9

800-572-1028                                            caltestelectronics.com                                                            8

3.5 Vertical Scale on Oscilloscope

The actual vertical scale of the oscilloscope is equal to the attenuation 

factor multiplied by the range of vertical scale selected on the 

oscilloscope. For example, using the CT3684, attenuation of 10x, the 

oscilloscope on 0.5 V/div, the real vertical scale is 10 x 0.5 = 5 V/div. 

With the CT3685, attenuation of 100x, the real vertical scale is 100 

x 0.5 = 50 V/div. These values apply when the oscilloscope is set to 

the typical 1 MΩ impedance input. When the oscilloscope is set to 50 

Ω input, the actual vertical scale will be doubled: 10 V/div for the 10x 

probe (CT3684) and 100 V/div for the 100x probe (CT3685). See Table 

1.

Vertical Scale on Oscilloscope

Scope Input 

Impedance

Probe 

Attenuation 

Setting

Actual 

Attenuation 

Setting

Vertical Scale 

Reading on the 

Oscilloscope

Actual Vertical 

Scale of the 

Oscilloscope

1 MΩ

10x

10x

0.5 V/div

5 V/div

1 MΩ

100x

100x

0.5 V/div

50 V/div

50 Ω

10x

20x

0.5 V/div

10 V/div

50 Ω

100x

200x

0.5 V/div

100 V/div

Table 1  Oscilloscope Readings

4  Cleaning

This probe does not require any particular cleaning. If necessary, clean 

the case with a cloth slightly moistened with soapy water. Make sure 

the probe is completely dry before reconnecting it to an oscilloscope.

WARNING

Dry the probe thoroughly before attempting to make voltage 

measurements.

CAUTION 

Do not subject the probe to solvents or solvent fumes as these can 

cause deterioration of the probe body and cables.

Содержание CT3684

Страница 1: ...CT3684 CT3685 User Manual...

Страница 2: ...ot work alone when working with high voltages Inspect the Probe Inspect the probe and accessories for cracks and frayed or broken leads before each use If defects or damages are noted DO NOT USE the p...

Страница 3: ...for Critical Applications This probe is not authorized for use in contact with the human body or for use as a component in a life support device or system Do Not Substitute Parts Do not install subst...

Страница 4: ...or measurements in circuits of installation CAT III it should incorporate the use of fused test probes Compliance Statements Disposal of Old Electrical Electronic Equipment Applicable in the European...

Страница 5: ...on mode CT3684 z z Up to 1400 V DC AC peak common mode CT3685 z z Attenuation settings of 10x CT3684 and 100x CT3685 z z High accuracy 1 z z High CMRR z z Power indicator LED z z 9 VDC adapter CT3723...

Страница 6: ...uct Description 2 1 CT3684 CT3685 Description Output BNC cable 90 cm with 4 mm Banana Plug Ground Lead Hook Probes Probe Body Power On Off Switch LED Power Indicator Leads 50 cm Figure 1 Front Panel D...

Страница 7: ...Optional Battery Pack If using the optional battery pack CT3729 the batteries supplied must be inserted into the battery pack and the pack attached to the probe See Figure 2 Figure 2 Replacing the Bat...

Страница 8: ...example 7 A wave matching the test signal should display on the screen of the oscilloscope and this means this probe is working properly 3 4 Getting Started 1 Connect the hook probes to the leads 2 C...

Страница 9: ...84 and 100 V div for the 100x probe CT3685 See Table 1 Vertical Scale on Oscilloscope Scope Input Impedance Probe Attenuation Setting Actual Attenuation Setting Vertical Scale Reading on the Oscillosc...

Страница 10: ...ge DC AC peak 70 V 700 V Maximum Common Mode Input Voltage DC AC peak 700 V 1400 V Absolute Maximum Rated Input Voltage each side to ground 500 Vrms CAT I 1000 Vrms CAT II Input Impedance 1 6 M 7 pF e...

Страница 11: ...s IEC 61010 031 CAT I CT3684 CATII CT3685 Specifications are subject to change without notice To ensure the most current version of this manual please download the current version from our website cal...

Страница 12: ...ersons will not be covered by this warranty Cal Test Electronics reserves the right to discontinue models change specifications price or design of this device at any time without notice and without in...

Страница 13: ...800 572 1028 caltestelectronics com 12...

Страница 14: ...800 572 1028 caltestelectronics com 13...

Страница 15: ...800 572 1028 caltestelectronics com 14...

Страница 16: ...2016 Cal Test Electronics Part 450844 001 20161222...

Отзывы: