SERVICE MANUAL
CALIBRATION
3-13
Set the TW’s frequency to 160 Hz with the following GPIB command:
SOUR:FREQ 160.0
Wait for the DMM reading to settle and record the value as
RDNG3
.
Query the TW’s output put voltage with the GPIB string below and record this value as
MEAS3
.
MEAS
{1,2,3}
:VOLT?
Set the TW’s frequency to 320 Hz with the following GPIB command:
SOUR:FREQ 320.0
Wait for the DMM reading to settle and record the value as
RDNG4
.
Query the TW’s output put voltage with the GPIB string below and record this value as
MEAS4
.
MEAS
{1,2,3}
:VOLT?
Set the TW’s frequency to 500 Hz with the following GPIB command:
SOUR:FREQ 500.0
Wait for the DMM reading to settle and record the value as
RDNG5
.
Query the TW’s output put voltage with the GPIB string below and record this value as
MEAS5
.
MEAS
{1,2,3}
:VOLT?
Set the TW’s frequency to 60 Hz and open the output with the following GPIB
commands:
SOUR:FREQ 60.0
OUTP OFF
Calculate the measurement system frequency compensation values:
VALUE1
=
RDGN1
/
MEAS1
VALUE2
=
RDGN2
/
MEAS2
VALUE3
=
RDGN3
/
MEAS3
VALUE4
=
RDGN4
/
MEAS4
VALUE5
=
RDGN5
/
MEAS5
Update the TW’s calibration data with the following GPIB command:
CAL
{1,2,3}
:MEAS:LOC:LOR:VOLT:FREQCAL
VALUE1 VALUE2 VALUE3 VALUE4 VALUE5
Содержание TrueWave TW5250
Страница 2: ......
Страница 4: ...ii This page intentionally left blank...
Страница 10: ...TABLE OF CONTENTS TW SERIES viii This page intentionally left blank...
Страница 36: ...MAINTENANCE AND TROUBLESHOOTING TW SERIES 2 16 This page intentionally left blank...
Страница 80: ...CALIBRATION TW SERIES 3 44 This page intentionally left blank...