SERVICE MANUAL
CALIBRATION
3-5
Send the following query command to the TW and record the return string as
MEAS1
:
MEAS
{1,2,3}
:VOLT?
Calculate the offset value:
VALUE1 = RDGN1 – MEAS1
Send the offset value to the TW:
CAL
{1,2,3}
:MEAS:LOC:LOR:VOLT:OFFSET
VALUE1
Measure the current offset by sending the following GPIB query and record the return
string as
MEAS1
:
MEAS
{1,2,3}
:CURR?
Calculate the offset value:
VALUE1
= -1.0 *
MEAS1
Send the calculated offset to the TW:
CAL
{1,2,3}
:MEAS:LOC:LOR:CURR:OFFSET
VALUE1
Set the TW to 220 volts DC by sending the following GPIB string:
SOUR
{1,2,3}
:VOLT:OFFSET 220
Wait for the DMM to settle. Take a DMM reading and record the value as
RDGN1
.
Query the output voltage of the TW and record the return string as
MEAS1
:
MEAS
{1,2,3}
:VOLT?
Set the TW to -220 volts DC by sending the following GPIB string:
SOUR
{1,2,3}
:VOLT:OFFSET –220
Wait for the DMM to settle. Take a DMM reading and record the value as
RDGN2
.
Query the output voltage of the TW and record the return string as
MEAS2
:
MEAS
{1,2,3}
:VOLT?