background image

NewLife

®

 Intensity 10

User Manual 

(English)

Содержание NewLife Intensity 10

Страница 1: ...NewLife Intensity 10 User Manual English...

Страница 2: ...Type BF applied part degree of protection against electric shock Reg 5333 Warning Reg W001 Council Directive 93 42 EEC concerning medical devices Authorized representative in the European Community I...

Страница 3: ...home when traveling or while participating in daily activities away from home Oxygen is a gas that makes up 21 of the room air we breathe Our bodies depend on a steady supply to function properly You...

Страница 4: ...n operate at any location where AC power source is available Oxygen concentrator use requires a physician s prescription and is not intended for life support use Although oxygen therapy can be prescri...

Страница 5: ...perating Instructions Explains procedures to operate the unit H Air Intake Gross Particle Filter Prevents dust and other airborne particles from entering the unit I Power Cord Allows connection of uni...

Страница 6: ...RDOUS THIS DEVICE SHOULD ONLY BE USED WHEN PRESCRIBED BY A PHYSICIAN WARNING NOT FOR USE IN THE PRESENCE OF FLAMMABLE ANESTHETICS WARNING AS WITH ANY ELECTRICALLY POWERED DEVICE THE USER MAY EXPERIENC...

Страница 7: ...ntal by order of a physician or other licensed health care provider CAUTION Do not position the unit so that it is difficult to access the power cord CAUTION The concentrator should be located as to a...

Страница 8: ...m capacity of the concentrator Note To prevent a void warranty follow all manufac turers instructions Note Do not attempt any maintenance other than the possible solutions listed within the manual Not...

Страница 9: ...Pressure 20 psig 138 kPA Operating Environment 41 F to 104 F 5 C to 40 C 15 90 humidity Altitude 1250 to 10 000 ft 381 to 3048 m tested to 700 1060 hPa Storage Environment 25 C to 70 C 13 F to 158 F...

Страница 10: ...if required nasal cannula face mask catheter and or extension tubing to the oxygen outlet 5 Completely unwrap the power cord 6 Insert power cord into the electrical outlet 7 Locate the power switch on...

Страница 11: ...BE USED OXYGEN AC CELERATES THE COMBUSTION OF FLAMMABLE SUBSTANCES WARNING THE USE OF SOME OXYGEN ADMIN ISTRATION ACCESSORIES NOT SPECIFIED FOR USE WITH THIS OXYGEN CONCENTRATOR MAY IMPAIR ITS PERFOR...

Страница 12: ...If your physician did not prescribe a humidi fier connect the oxygen tubing directly to the unit s oxygen outlet A separate outlet fitting is supplied for this type of connection Operating With Humidi...

Страница 13: ...pter or humidifier Note To Equipment Provider The following oxygen administration accessories are recommended for use with the NewLife Oxygen Concentrator Nasal Cannula with 7 feet 2 1 m of tubing 6 L...

Страница 14: ...ermal switch located within the compressor trips shutting down the compressor Oxygen Monitor In the event the oxygen monitor detects an oxygen concentration below 82 the low oxygen concentration warni...

Страница 15: ...cation as prescribed by your physician Note Use only the amount of medication and frequen cy of treatment that your physician prescribed 3 Remove or unscrew the medication cup on the nebulizer and pl...

Страница 16: ...er parts 1 After each treatment separate the nebulizer and the T piece or mouthpiece assembly 2 Remove or unscrew the nebulizer cup and rinse each component thoroughly in warm water 3 Once a day clean...

Страница 17: ...ICAL SHOCK HAZARD TURN OFF THE UNIT AND DISCONNECT THE POWER CORD FROM THE ELECTRIC OUTLET BEFORE YOU CLEAN THE UNIT TO PREVENT ACCIDENTAL ELECTRICAL SHOCK AND BURN HAZARD ONLY YOUR EQUIP MENT PROVIDE...

Страница 18: ...least one time each week wash the air intake gross particle filter which is located in the back of the unit Your Equipment Provider may advise you to clean it more often depending upon your operating...

Страница 19: ...d intermittent audible alarm Power Failure Ensure device is plugged into a known working outlet Ensure breaker switch is pushed in If issue persists contact equipment provider for service Troubleshoot...

Страница 20: ...e unit is positioned away from curtains or drapes hot air registers heaters and fireplac es Be certain to place the unit so all sides are at least 12 inches 30 5 cm away from a wall or other obstructi...

Страница 21: ...ard Accessories Nasal Cannula with 7 feet 2 1 m of tubing 6 LPM max CU002 1 Oxygen Outlet Adapter F0025 1 Face Mask with 7 feet 2 1 m of tubing 10 LPM max MS013 1 Humidifier Adapter Extension HU002 1...

Страница 22: ...ed in such an environment Emissions Test Compliance Electromagnetic Environment Guidance RF emissions CISPR 11 Group 1 The NewLife uses RF energy only for its internal function There fore its RF emiss...

Страница 23: ...l on AC lines 2kV common mode on outdoor I O lines N A 1kV differential on AC lines N A Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interrupti...

Страница 24: ...lectromag netic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz a...

Страница 25: ...distance between the transmitting antenna and the ME EQUIPMENT OR ME SYSTEM may be reduced to 1m The 1 m test distance is permitted by IEC 61000 4 3 a For some services only uplink frequencies are in...

Страница 26: ...the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply i...

Страница 27: ...PN MN239 F User Manual ENG 27 NewLife Intensity 10 Notes...

Страница 28: ...IRE Inc Please visit our website below for a full listing of trademarks Trademarks www caireinc com corporate trademarks Copyright 2022 CAIRE Inc All Rights Reserved CAIRE Inc reserves the right to di...

Отзывы: