MN240-1 Rev C
33
5.8
Product Tank Replacement
Product Tank Removal
To remove the product tank, follow the steps below:
1.
Set the unit’s ON/OFF power switch to the OFF position and disconnect the
power cord.
2.
Remove the sides, lower front, and back panels.
3.
Remove the compressor. (Refer to Section 5.3.)
4.
Remove the intake resonator, which is held in place by Velcro.
5.
Remove the securing nut that holds the product tank to the superstructure. (This
securing nut is located in the top right area of the compressor compartment).
6.
Remove the screws that hold the control panel to the superstructure. It is not
necessary to fully remove the control panel. It is only necessary to create enough
room to allow the product tank to be removed from the unit.
7.
Remove the green tubing located at the regulator outlet and at the bottom of the
product tank assembly.
8.
Cut the tie wrap securing the product tank to the superstructure.
9.
Lift the product tank with regulator attached from the mounting hole and remove
from the unit.
10.
Carefully remove regulator from product tank.
Product Tank Installation
To install the product tank assembly, follow the steps below:
1.
Perform the product tank removal procedure in reverse order.
2.
Apply sealant or sealant tape to threads of product tank fitting before reinstalling
regulator onto product tank.
3.
Perform the procedure from Section 5.9.1 to set the regulator to 20 PSIG (138
kPa). Leak test all connections.
Содержание NewLife Elite
Страница 1: ...MN240 1 Rev C 1 NewLife Family Technical Manual MN240 1...
Страница 6: ...MN240 1 Rev C 6...
Страница 7: ...MN240 1 Rev C 7...
Страница 8: ...MN240 1 Rev C 8...
Страница 9: ...MN240 1 Rev C 9...
Страница 10: ...MN240 1 Rev C 10...
Страница 11: ...MN240 1 Rev C 11...
Страница 46: ...MN240 1 Rev C 46...
Страница 47: ...MN240 1 Rev C 47...
Страница 48: ...MN240 1 Rev C 48...