MN240-1 Rev C
16
Product Specifications Continued
5 Liter Concentrators
8 Liter Concentrators
10 Liter Concentrators
Backpressure
Effect Test
Flow
Applied Pressure *
5.00 LPM
@ 0 PSIG (0 kPA)
4.75 LPM
@ 1 PSIG (7 kPA)
Flow
Applied Pressure *
8.00 LPM @ 0 PSIG (0 kPA)
7.86 LPM @ 1 PSIG (7 kPA)
Flow
Applied Pressure *
10.00 LPM @ 0 PSIG (0 kPA)
9.65 LPM
@ 1 PSIG (7 kPA)
Outlet
Pressure Test
Actual
Pressure
Max Allowable
Limit
7.3 PSIG
(50.3 kPA)
8.8 PSIG
(60.7 kPA)
Actual
Pressure
Max Allowable
Limit
21.9 PSIG
(151 kPA)
22.0 PSIG
(151.7 kPA)
Actual
Pressure
Max Allowable
Limit
21.9 PSIG
(151 kPA)
22.0 PSIG
(151.7 kPA)
(1)
Based on 21
o
C (70°F) at a nominal operating pressure range of 0-7 kPa (0-1 PSIG back pressure)
and 14.7 PSI (101 kPa).
(2)
Operating outside of these operational specifications can
affect performance.
Accuracy: ±10% of indicated flow setting or ±200 mL, whichever is greater. Based on an atmospheric
pressure range of 700 kPa to 1060 kPa at 70°F (21°C).