HELiOS Portables Technical Service Manual • PN 20562190 Rev E
18
HELiOS P
or
tables
Operation
IX
PN 20562190 Rev A
Operation (Using the Oxygen Supply Line)
1. Verify that there is adequate liquid oxygen in the reservoir.
2. Insert the flexible oxygen supply tube connector (P/N
B-701656-SV) into the quick connect on the front of the
H300 or H850 and snap it into place.
3. Locate the DISS nut and tailpiece assembly attached to the
opposite end of the flexible oxygen supply tube.
4. Thread the nut and tailpiece assembly onto the HELiOS res-
ervoir oxygen outlet connector until secure.
5. Push one of the tubes from the dual lumen cannula onto the
top cannula barb on the H300 or H850. Place the other tube
onto the lower cannula barb.
6. Properly position the nasal cannula on the patient.
7. Turn the flow control knob on the top of the unit to the pre-
scribed demand flow setting.
NOTE: If a hissing sound is heard, tighten the connection until
the hissing stops.
Figure 18:
HELiOS Supply Line Connection
WARNING:
Use only the recommended flexible oxygen
supply tube and oxygen gas source. Connecting the unit to
an incorrect gas source can cause inhalation of hazardous
substances.
NOTE: The H850 must be used in demand (D) flow settings
when operating from the oxygen supply line.
CAUTION: Connect the portable only to a reservoir or other
gaseous oxygen source that operates at pressures between
1.41 to 1.62 bar (20.5 and 23.5 psi). Oxygen sources outside
of this range can cause the unit to operate improperly or
damage the unit.