HELiOS Portables Technical Service Manual • PN 20562190 Rev E
16
HELiOS P
or
tables
Unpacking and Setup
VIII
Unpacking
1. Inspect carton for shipping damage. Report any damage
to freight company before signing bill of lading.
2. Check description on carton against your order.
3. Unpack unit, including all accessories and documenta-
tion.
4. Set aside packing materials in case unit must be returned
to the factory.
Setup
1. Record the serial number of the unit.
2. Visually inspect the unit for damage from improper han-
dling. Note any dents, cracks, or missing parts.
3. Ensure the vent valve is in the closed position. Be sure
its lever opens and closes smoothly.
4. If possible, connect the H300 or H850 to a reservoir to
check for smooth coupling.
5. Hold the unit on a table and carefully pull up on the con-
tents indicator strap. Verify that the mechanism moves
smoothly.
6. Verify that the unit has all its warning/caution labels.
Check labels for damage.
7. Wipe away any dust on the unit with a clean, dry, lint-
free cloth.
8. Verify that a standard dual lumen cannula will attach to
the unit. It should be rated at least 6 LPM in flow and
should be a maximum of 7 ft (213.4 mm) in length.