Personal Ambulatory Oxygen System
Provider Technical Manual
78
PN 20631679 Rev D
NOTE:
There are no field serviceable parts inside the Compressor Box. Do not
attempt to disassemble or modify the Compressor Box in the field.
CAUTION
The Eclipse 5 contains electrostatic sensitive components. Do not open
or handle except at a static free workstation.
Pry
Point
Pry
Point
Pry
Point
Case Bottom
Cover
Compressor
As with any concentrator, the compressor is a limited-wear component and may require servicing
during the lifetime of the device. The point of service will be dependent on factors such as operation
time, flow settings and environmental conditions. Service is required when oxygen purity and/or flow
rates cannot be maintained.
REMOVE AND REPLACE THE COMPRESSOR BOx
1. Remove the Unit Cover as described in the section Remove and Replace the Unit Cover.
2. Flip unit over onto a padded ESD safe surface. Using a small flat tip screwdriver to pry open the cover at each
pry point, remove the Case Bottom Cover as shown in Figure 55.
Figure 55: Case Bottom Cover Removal