Document type:
Title:
Revision date:
Revision:
User's Manual (MUT)
Mod. N568B/LC 16 Ch Spectroscopy Amplifiers
26/01/2007
4
Filename:
Number of pages:
Page:
N568B_N568LC_REV4.DOC
37
17
AA Ch** Fg NNN
N568 Cg M
where AA is still the selected address number. The address number starts blinking, indicating the
currently modifiable field. By pressing the key "#", the blinking can be moved on the Ch (Channel
Number) field: by pressing "ROLL" it is possible to increment the value of the Channel Number by 1
(Channel values: 0 to 15). This number indicates the Channel on which all following settings are
performed. By pressing the key "#" the blinking toggles to the Fg (Fine Gain) field in the display (NNN
in the figure above): by pressing "0" or "1" it is possible respectively to decrement/increment the value
of the Fine Gain by 1, while by pressing "4" or "7" it is possible respectively to decrement/increment
the value of the Fine Gain by 10 (allowed Fine Gain values: 0 to 255). By pressing the key "#" the
blinking toggles to the Cg (Coarse Gain) field in the display (M in the figure above): by pressing
"ROLL" it is possible to increment the value of the Coarse Gain by 1 (allowed Coarse Gain values: 0
to 7).
By pressing the "PAGE" key, the display will appear as follows:
AA Ch** PZ NNN
N568 Sh L
where AA is the selected address number. By pressing the key "#" the blinking toggles from the
position of the blinking field in the previous page to the following at its right. By pressing the key "#",
the blinking can be moved on the Ch (Channel Number) field: by pressing "ROLL" it is possible to
increment the value of the Channel Number by 1 (Channel values: 0 to 15). This number indicates the
Channel on which all following settings are performed. By pressing the key "#" the blinking toggles to
the PZ (Pole Zero adjustment) field in the display (NNN in the figure above): by pressing "0" or "1" it is
possible respectively to decrement/increment the value of the Pole Zero adj. by 1, while by pressing
"4" or "7" it is possible respectively to decrement/increment the value of the Pole Zero adj. by 10
(allowed Pole Zero adj. values: 0 to 255). By pressing the key "#" the blinking toggles to the Sh
(Shaping Time) field in the display (L in the figure above): by pressing "ROLL" it is possible to
increment the value of the Shaping Time by 1 (allowed Shaping Time values: 0 to 3).
By pressing the "PAGE" key, the display will appear as follows:
AA Ch** Out Pos
N568 Out NInv
where AA is the selected address number. By pressing the key "#" the blinking toggles from the
position of the blinking field in the previous page to the following at its right. By pressing the key "#",
the blinking can be moved on the Ch (Channel Number) field: by pressing "ROLL" it is possible to
increment the value of the Channel Number by 1 (Channel values: 0 to 15). This number indicates the
Channel on which all following settings are performed. By pressing the key "#" the blinking toggles to
the higher row OUT (Output Polarity) field in the display (Pos in the figure above): by pressing "ROLL"
it is possible to switch from Positive (Pos) to Negative (Neg) polarity. By pressing the key "#" the
blinking toggles to the lower row OUT (Output configuration) field in the display (Ninv in the figure
above): by pressing "ROLL" it is possible to switch from Non-inverting (Ninv) to Inverting (Inv)
configuration.