16
$CMD:MON,CH:XX,PAR:IMIN
Min value of ISET
$CMD:MON,CH:XX,PAR:ISDEC
Decimal digits of ISET
$CMD:MON,CH:XX,PAR:IMON
IMON value
$CMD:MON,CH:XX,PAR:IMRANGE
IMON range (high /low)
$CMD:MON,CH:XX,PAR:IMRESL
IMON resolution in low range
$CMD:MON,CH:XX,PAR:IMRESH
IMON resolution in high range
$CMD:MON,CH:XX,PAR:IMDECL
Decimal digits of IMON in low range
$CMD:MON,CH:XX,PAR:IMDECH
Decimal digits of IMON in high range
$CMD:MON,CH:XX,PAR:MAXV
VMAX value
$CMD:MON,CH:XX,PAR:MVMIN
VMAX minimum value
$CMD:MON,CH:XX,PAR:MVMAX
VMAX maximum value
$CMD:MON,CH:XX,PAR:MVRES
VMAX resolution
$CMD:MON,CH:XX,PAR:MVDEC
Decimal digits of VMAX
$CMD:MON,CH:XX,PAR:PDWN
Power down mode Ramp / Kill
$CMD:MON,CH:XX,PAR:POL
Polarity
$CMD:MON,CH:XX,PAR:STAT
Status
IS_ON
1
Channel On
IS_UP
2
Channel Ramping Up
IS_DOWN
4
Channel Ramping Down
IS_OVC
8
Channel in Overcurrent
IS_OVV
0x10
Channel in OverVoltage
IS_UNV
0x20
Channel in UnderVoltage
IS_MAXV
0x40
Channel in MaxV
IS_TRIP
0x80
Channel in Trip
IS_MAXPW
0x100
Channel in MaxPower
IS_TWARN
0x200
Channel in Temperature warning (>80°C)
IS_OVT
0x400
Channel in OVT (>125°C)
IS_KILL
0x800
Channel Killed
IS_INTLCK
0x1000
Channel in Interlock
$CMD:MON,CH:XX,PAR:RUP
RAMP UP value
$CMD:MON,CH:XX,PAR:RUPMIN
RAMP UP minimum value
$CMD:MON,CH:XX,PAR:RUPMAX
RAMP UP maximum value
$CMD:MON,CH:XX,PAR:RUPRES
RAMP UP resolution
$CMD:MON,CH:XX,PAR:RUPDEC
Decimal digits of RAMP UP
$CMD:MON,CH:XX,PAR:RDW
RAMP DOWN value
$CMD:MON,CH:XX,PAR:RDWMIN
RAMP DOWN minimum value
$CMD:MON,CH:XXPAR:RDWMAX
RAMP DOWN maximum value
$CMD:MON,CH:XX,PAR:RDWRES
RAMP DOWN resolution
$CMD:MON,CH:XX,PAR:RDWDEC
Decimal digits of RAMP DOWN
$CMD:MON,CH:XXPAR:TRIP
Trip value
$CMD:MON,CH:XX,PAR:TRIPMIN
Trip minimum value
$CMD:MON,CH:XX,PAR:TRIPMAX
Trip maximum value
$CMD:MON,CH:XX,PAR:TRIPRES
Trip resolution
$CMD:MON,CH:XX,PAR:TRIPDEC
Decimal digits of Trip
Set Commands
$CMD:SET,CH:XX,PAR:VSET,VAL:XXX.XX
VSET value
$CMD:SET,CH:XX,PAR:ISET,VAL:XXXX.XX
ISET value
$CMD:SET,CH:XX,PAR:IMRANGE,VAL:LOW
IMON range low
$CMD:SET,CH:XX,PAR:IMRANGE,VAL:HIGH
IMON range high
$CMD:SET,CH:XX,PAR:RUP,VAL:XXX
RAMP UP value
$CMD:SET,CH:XX,PAR:RDW,VAL:XXX
RAMP DOWN value
$CMD:SET,CH:XX,PAR:PDWN,VAL:RAMP
POWER DOWN ramp mode
$CMD:SET,CH:XX,PAR:PDWN,VAL:KILL
POWER DOWN kill mode
$CMD:SET,CH:XX,PAR:ON
Set Ch ON
$CMD:SET,CH:XX,PAR:OFF
Set Ch OFF