CAEN
Electronic Instrumentation
DT55xxE Desktop HV Power Supply
13
Fig. 11: Terminal Ethernet settings
At first Power On the module is configured with default static IP (factory setting); such IP can be updated
using the “arrows” to select the fields, typing the new values and confirming with <Enter>.
Type S to save the new setting in the EEPROM and go back to Main Menu.
The new setting will become active at next Power On; if a DHCP Server is available, then the module can
be enabled or disabled as DHCP client; type S to save the new setting in the EEPROM and go back to Main
Menu: the DHCP server will automatically assign a new IP to the module at next Power On.
Firmware upgrade
To upgrade the firmware:
•
download from www.caen.it unit page the most recent firmware revision for your module
•
connect to the module via USB using Tera Term VT Emulator
•
in the Tera Term options, sele
ct “set up” > “serial port” a
nd enter the following settings
•
click OK to confirm
•
regarding Fig. 10: Terminal Board Menu, type U in order to upgrade the firmware
•
Type y
•
The following message will be shown:
•
Select
“
File
”
> disconnect
•
unplug the USB cable
•
do not turn off the module!
•
connect once again to the module, via USB
•
the following message will be shown:
•
Press any key
•
Wait until the following message is shown: