CAEN
Electronic
Instrumentation
DT54xx
USB
High
Voltage
Power
Supply
14
$CMD:MON,PAR:ISET<CR,LF>
Read
out
ISET
value
$CMD:MON,PAR:VMON<CR,LF>
Read
out
VMON
value
$CMD:MON,PAR:IMON<CR,LF>
Read
out
IMON
value
$CMD:MON,PAR:RUP<CR,LF>
Read
out
RUP
(RAMP
UP)
value
$CMD:MON,PAR:RDW<CR,LF>
Read
out
RDW
(RAMP
DOWN)
value
$CMD:MON,PAR:PDWN<CR,LF>
Contains
'RAMP'
or
'KILL'
$CMD:MON,PAR:IMRANGE<CR,LF>
Contains
'HIGH'
or
'LOW'
$CMD:MON,PAR:TRIP<CR,LF>
Read
out
TRIP
value
$CMD:MON,PAR:MAXV<CR,LF>
Read
out
Vmax
hardware
value
Vmax
can
be
set
via
hardware.
$CMD:MON,PAR:IMAX<CR,LF>
Read
out
ISET
maximum
value
$CMD:MON,PAR:STAT<CR,LF>
Read
out
Status
word
Bit
0
→
ON
1
:
ON
0
:
OFF
Bit
1
→
RUP
1
:
Channel
Ramping UP
Bit
2
→
RDW
1
:
Channel
Ramping DOWN
Bit
3
→
OVC
1
:
Over
current
Bit
4
→
OVV
1
:
Over
voltage
Bit
5
→
UNV
1
:
Under
voltage
Bit
6
→
MAXV
1
:
VOUT
in
MAXV
protection
Bit
7
→
TRIP
1
:
Current
generator
Bit
8
→
OVP
1
:
Over
temperature
Bit
9
reserved
Bit
10
→
DIS
1
:
Ch
disabled
Bit
11
→
KILL
1
:
Ch
in
KILL
Bit
12
→
ILK
1
:
Ch
in
INTERLOCK
Bit
13
→
NOCAL
1
:
Calibration
Error
Bit
14,
15
→
N.C.
reserved
Set
Commands
$CMD:SET,PAR:VSET,VAL:xxx.xx<CR,LF>
Set
VSET
value
$CMD:SET,PAR:ISET,VAL:xxx.xx<CR,LF>
Set
ISET
value