CAEN

 

 

Electronic Instrumentation

 

 

A7030-AG7030 3kV/ 1mA (1.5W) HV Boards 

11 

E-TRIPPED 

channel OFF due to external TRIP line signal

7

 

I-TRIPPED 

channel OFF due to internal OVERCURRENT condition 

EXT_DIS 

channel disabled by board INTERLOCK protection 

PWR_FAIL 

channel OFF due to exceeded power limit (>1.5W) 

After a E-TRIPPED, I-TRIPPED, EXT_DIS notification, it is necessary to perform a CLEAR ALARM cycle, before turning 
the channel ON.

 

Moreover it is possible to monitor board parameters, such as measured Temperature, HIMax and HVMax, 
and to check board status; the following messages may be returned by the POWER SUPPLY SYSTEM when 
monitoring the board status: 
UNDER_TEMP     

board temperature < 5°C  

OVER_TEMP       

board temperature > 65°C 

Interlock protection 

 

 

Fig. 2 

 INTERLOCK electrical scheme 

In order to enable the HV output channels, first of all it is necessary that SAFETY LOOP pin on the Radiall 
52pin output connector is short circuited with the next RETURN pin (pin 21 on A/AG7030 

 7030T; pin 10 

on A/AG7030L); if the board features SHV connectors, skip this step. 

Then it is necessary to configure the 

Interlock connector as follows:

 

contact open 

INTERLOCK 

voltage level (0÷1V, ~5mA current) between pin 2 and pin 3 

INTERLOCK 

short circuit pin 1 with pin 2, and pin 3  with pin 4 

ENABLED 

voltage level (4÷6V, ~5mA current) between pin 2 and pin 3 

ENABLED 

A schematic diagram of the Interlock input is shown in the figure above, where the diode is part of opto-coupler stage. 

Interlock

 means that channel is hardware disabled. 

The front panel Interlock LED is ON when the INTERLOCK is active;  as INTERLOCK is active, channels are turned off at 
the fastest available rate, regardless the RAMP DOWN setting. 

Grounding specifications 

The Mod. A7030 channels share a common floating return (HVGND),  insulated from the crate ground (DGND). This 
feature allows on-detector grounding, thus avoiding loops which may increase noise level. HVGND and DGND may be 
connected, by short circuiting J1 jumper pins on the motherboard (see figure below). 

 

Fig. 3 

 J1 jumper location 

 

 
 

7

 EXTTRIP and INTTRIP parameters are expressed in Hexadecimal format 

Содержание A7030

Страница 1: ...CAEN Tools for Discovery n Electronic Instrumentation Rev 5 Es A7030 AG7030 3kV 1mA 1 5W HV Boards Rev 7 03 September 2019...

Страница 2: ...t is defective due to workmanship or materials and has not been caused by mishandling negligence on behalf of the User accident or any abnormal conditions or operations CAEN declines all responsibilit...

Страница 3: ...cautions 9 Safety Terms and Symbols on the Product 9 Installation 9 3 Operating modes 10 Output control and monitoring 10 Channel STATUS Flag 10 Interlock protection 11 Grounding specifications 11 Saf...

Страница 4: ...red via software if exceeded the relevant channel can be programmed either to turn off after a programmable trip time or to keep on providing the maximum allowed current this particular feature allows...

Страница 5: ...esolution 1 V Ramp Up Ramp Down 1 500 Volt sec 1 Volt sec step Voltage Ripple 2 10 1000 Hz 15 mV typical 20 mV max 1000 Hz 5 mV typical 10 mV max VMon vs VOut Accuracy 3 4 typical 0 3 0 2 V max 0 3 1...

Страница 6: ...CAEN Electronic Instrumentation A7030 AG7030 3kV 1mA 1 5W HV Boards 6 Front Panel Fig 1 A7030 family front panel 48 36 24 Multipin 24 SHV and 12 SHV channel version...

Страница 7: ...ge VMAX common to all the channels Its value can be read out via software IMAX Function it allows to adjust the hardware maximum current IMAX common to all the channels Its value can be read out via s...

Страница 8: ...CH26 37 CH23 48 CH25 7 CH32 17 CH29 28 CH31 38 CH28 49 CH30 8 CH37 18 CH34 29 CH36 39 CH33 50 CH35 9 CH42 19 CH39 30 CH41 40 CH38 51 CH40 10 CH47 20 CH44 31 CH46 41 CH43 52 CH45 21 RETURN 42 SAFETY L...

Страница 9: ...rs To avoid electric shock or fire hazard do not operate this product with covers or panels removed Do Not Operate in Wet Damp Conditions To avoid electric shock do not operate this product in wet or...

Страница 10: ...POn EN DIS Power On option which can be enabled or disabled If this option is enabled at Power On or after a Restart each channel is restored in the same condition defined by the Power parameter it wa...

Страница 11: ...pin pin 21 on A AG7030 7030T pin 10 on A AG7030L if the board features SHV connectors skip this step Then it is necessary to configure the Interlock connector as follows contact open INTERLOCK voltage...

Страница 12: ...not a working one The Connector Configurator allows to optimize the connection of the return and of DGND Earth The best configuration must be determined by the user upon application the optimal conne...

Страница 13: ...ustomers including engineers scientists and technical professionals who all trust them to help achieve their goals faster and more effectively CAEN S p A CAEN GmbH CAEN Technologies Inc Via Vetraia 11...

Отзывы: