Calibration Procedure
C
ALIBRATION
P
ROCEDURE
VimedixAR supports the following two methods to calibrate the position of the Hologram manikin
relative to the physical manikin:
• Calibration by Tag
• Calibration by Gaze
Note:
Calibration by Tag or Gaze can always be repeated if the physical manikin is moved.
Regardless of which Shared Experience Mode is selected or which form of calibration is used, access
to the following buttons during calibration are enabled:
Information
,
Calibration
,
Adjust Menu
,
and
Help
. All other buttons including
Hologram Controls
are disabled during calibration.
Calibration by Tag
To start Calibration by Tag:
1. On the Interactive Panel, select the
Calibration
button.
2. Select the
Use Tag
button at the bottom of the Calibration menu.
Once the
Use Tag
button is selected, a small target symbol appears at the top left corner of the field
of view with the message “Place the hologram using the tag.” Now place the provided “T” button on
the table below the right ear of the Physical manikin with the arrow facing the head. When the
HoloLens 2 recognizes the tag, it places a holographic image of the tag on top of it with the Hologram
manikin not far from it. You must then move the tag to change the position of the Hologram manikin
to match the Physical manikin.
Select the
End
button at the bottom of the Interactive Panel to end the calibration by tag.
Calibration by Gaze
To start a Calibration by Gaze:
a. On the Interactive Panel, select the
Calibration
button
b. Select the
Use Gaze
button at the bottom of the Calibration menu.
One the
Use Gaze
button is selected, a small target symbol appears at the top left corner of the field
of view with the message, “Air Tap to place the hologram.” With your Gaze, move the Hologram
manikin to match the position and Air Tap the Hologram manikin to anchor it in place.
Select the
End
button at the bottom of the Interactive Panel to end the calibration by Gaze.
Содержание VimedixAR
Страница 1: ...Your worldwide training partner of choice User Guide with HoloLens 2...
Страница 6: ...Contents iv 2020 CAE 905K803052 CAEVimedixAR THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...Cautions and Warnings 4 2020 CAE 905K803052 CAEVimedixAR THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 12: ...Helpful Links 6 2020 CAE 905K803052 CAEVimedixAR THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 14: ...Equipment Overview 8 2020 CAE 905K803052 CAEVimedixAR THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 48: ...CAE VIMEDIXAR User Interface UI 44 2020 CAE 905K803052 CAEVimedixAR THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 52: ...Calibration Procedure 48 2020 CAE 905K803052 CAEVimedixAR THIS PAGE INTENTIONALLY LEFT BLANK...