background image

Customer Care Australia P: 133 122 F: 1300 303 310 E: [email protected] W: cabac.com.au 

A: 8 Distribution Pl, Seven Hills NSW 2147
Customer Care New Zealand P: 8 0077 6969 F: 8 0088 6969 E: [email protected] W: cabac.co.nz

8

 

 

 

4.9 MEASURING RESISTANCE

 

4.9.1

 

Rotate the probe socket clockwise to spin out the probe from the meter.

 

4

.9.2 Insert the black test lead or test clip in the 

COM

 

jack

 

.

 

4.9.3 Set the transform switch at the 

Ω

 

range position

 

Auto range or manual range can be

 

transformed

 

 

by putting the “

RANGE

 b

utton

4.9.4 Connect the probe tip of the

 

 

meter to one side of the

 

tested

 

 

circuit and probe tip of

 

the test

 

 

lead (or test clip) to the other side.

 

4.9.5 You can get reading from LCD

 

display

 

  WARNING 

Risk of Electrocution.

 

When measuring in

-

circuit resistance, be sure the circuit under test has all power removed and 

that all capacitors have been discharged fully.

 

 

NOTE:

 

At the manual range mode, when only the figure ‘OL’ is displayed, it indicates over

-

range situation and the higher range 

has to be selected.

 

For measuring resistance above 1MΩ, the meter may take a few seconds to get stable reading.

 

   

When the input is not connected, i.e. at open circuit, the figure ‘OL’ will be displayed for the over

-

range condition.

 

 

4.10

 

TESTING DIODE

 

4.10.1

 

Rotate the probe socket clockwise to spin out the probe from the meter.

 

4.10.

Insert the black test lead or test clip in the 

COM

 

jack.

 

4.10.

Set the transform switch at the 

 

range position.

 

4.10.

Put down the "

FUNC.

" button

 

transformed at 

 

test. 

 

4.10.

Connect the probe tip of the meter to the anode of the diode, 

 

   

the black test lead (or test clip) to the cathode of the diode.

 

4.10.

You can get reading from LCD

 

display

 

NOTE:

 

The meter will show the 

 

approximate forward voltage

 

drop of the diode. 

 

   

If the meter and the lead connection is reversed, only figure ‘OL’ will be displayed.

 

When the input is not connected, i.e. at open circuit, the figure ‘OL’ will be displayed.

 

 

4.11

 

CONTINUITY TEST

 

  WARNING 

Risk of Electrocution.

 

When testing the circuit continuity, be sure that the power of

 

the circuit has been shut down and 

all capacitors have been discharged fully.

 

 

4.11.1

 

Rotate the probe socket clockwise to spin out the probe from the meter.

 

4.11.

Insert the black test lead or test clip in the 

COM

 

jack.

 

4.11.

Set the transform switch

 

  

at the 

 

range position.

 

4.11.

Put down the "

FUNC.

 

  

button

 

transformed at 

  

  

continuity test. 

 

4.11.

Connect the probe tip of 

 

   

the meter to one side of the

 

             

 

tested circuit and probe tip of the test lead (or test clip) to the other side.

 

4.11.

If continuity exists 

(

i.e., resistance less than about 50Ω

)

, built

-

in buzzer will sound.

 

 

NOTE:

 

If the input open circuit (or the circuit resistance measured is higher than 200Ω), then the figure‘0L’ will be displayed.

 

 
 
 
 
 

Содержание T8211D

Страница 1: ...1300 303 310 E sales cabac com au W cabac com au A 8 Distribution Pl Seven Hills NSW 2147 Customer Care New Zealand P 8 0077 6969 F 8 0088 6969 E sales cabac co nz W cabac co nz 1 PEN TYPE DIGITAL MULTIMETER OPERATION MANUAL T8211D T8211D ...

Страница 2: ... 1 1 3 When poor condition under harsh preservation or shipping conditions caused inspect and confirm this meter without delay 1 1 4 Test lead or test clip must be in good condition Before using verify that the insulation on test lead or test clip is not damaged and or the leads wire is not exposed 1 1 5 Full compliance with safety standards can be guaranteed only if used with test leads supplied ...

Страница 3: ...ect the tip of meter and the probe of the test lead or test clip from all tested circuits 1 4 3 To avoid the wrong reading causing electricity attack when the meter displays you must change the battery 1 4 4 Do not use abrasives or solvents on the meter use a damp cloth and mild detergent only 1 4 5 Always set the transform switch to the OFF position when the meter is not in use 1 4 6 If the meter...

Страница 4: ...D Display The measuring result will be displayed LED Indicator At Logic test range High level and Low level can be indicated Rotatable Probe Socket Rotating the probe in or out of the meter Protection Ring To keep the hand from the probe behind the ring 3 SPECIFICATIONS Accuracy is specified for a period of year after calibration and at 18 to 28 64 to 82 with relative humidity to 75 3 1 GENERAL SP...

Страница 5: ...0V 0 1V 600V 1V 1 0 of rdg 3 digits Input Impedance 10MΩ Overload Protection 200mV range 250V DC or AC rms 2V 600V ranges DC 600V or AC 600V rms Frequency Range 40 to 400Hz Response Average calibrated in rms of sine wave Max Input Voltage 600V rms AC 3 2 3 Resistance Range Resolution Accuracy 200Ω 0 1Ω 1 0 of rdg 3 digits 2kΩ 0 001kΩ 20kΩ 0 01kΩ 1 0 of rdg 1 digit 200kΩ 0 1kΩ 2MΩ 0 001MΩ 20MΩ 0 01...

Страница 6: ...ormed between DC and AC range Put FUNC button when measuring the resistance diode and continuity meter will transform among them 4 4 RANGE TRANSFORM The auto range is used when measuring the voltage and resistance Put down the RANGE button if the manual range is needed Each time you put down range will go upward the minimum range is transformed if RANGE button is put down at the maximum range If t...

Страница 7: ...ure OL is displayed it indicates over range situation and the higher range has to be selected At the manual range mode when the value scale to be measured is unknown beforehand select the range to the highest position and set down gradually 4 8 MEASURING AC VOLTAGE 4 8 1 Rotate the probe socket clockwise to spin out the probe from the meter 4 8 2 Insert the black test lead or test clip in the COM ...

Страница 8: ... probe socket clockwise to spin out the probe from the meter 4 10 2 Insert the black test lead or test clip in the COM jack 4 10 3 Set the transform switch at the range position 4 10 4 Put down the FUNC button transformed at test 4 10 5 Connect the probe tip of the meter to the anode of the diode the black test lead or test clip to the cathode of the diode 4 10 6 You can get reading from LCD displ...

Страница 9: ...spin out the probe from the meter 4 13 2 Insert the black test lead or test clip in the COM jack 4 13 3 Set the transform switch at the mA range position 4 13 4 Press the FUNC button to transform to AC measurement Auto range and manual range can be transformed by putting the RANGE button 4 13 5 Connect the probe tip of the meter and probe tip of the test lead or test clip in series with the load i...

Страница 10: ...move it 5 1 3 Replace the exhausted battery with a new one 5 1 4 Put the battery cover as its origin 5 2 TEST LEADS OR TEST CLIP REPLACEMENT WARNING Full in compliance with safety standards can be guaranteed only if used with test leads supplied If necessary they must be replaced with the same model or same electric ratings Electric ratings of the test leads 600V 10A You must be replaced the test ...

Отзывы: