MAINTENANCE
The
NEW ASKIR 30
suction equipment does not need maintenance or lubrication.
It is necessary to check functioning and instrument before every use.
Unpack the instrument and
always check
integrity of plastic parts and feeding cable, they might have been
damaged during previous use.
Connect cable to electrical network and turn switch on.
Close the aspiration outlet with your finger and with suction regulator in maximum vacuum position check
that the vacuum indicators reaches -80 kPa (-0.80 bar) maximum.
Rotate the knob from right to left and check the aspiration regulating control.
The vacuum indicator should go down -40 kPa (-0.40 bar).
Verify that loud noises are not present, these can indicate wrong functioning.
A protection fuse (
F 1x1.6A 250V and F 1 x 4A 250V for voltage 110V/60Hz
) reachable from exterior and it
situated in the plug protects the instrument. For use replacing, always check the type and the range
indicated.
Fault type
Cause
Solution
1. No aspiration
Jar Cap badly screwed
down
Unscrewed the cap, then rescrew it correctly
2. No aspiration
Lid seal not in its seat
Unscrew the cap and insert the seal properly
in its seat
3. The float doesn’t close
If the cap has been
washed, ensure that the
float is not partially
detached
Insert the float into it’s place
4. The float doesn’t close
The float it’s covered by
dirty material
Unscrewed the cap, leave the and put in on
autoclave
5. Low suction
Foam inside the jar
Fill the jar to 1/3 full of ordinary water
6. No aspiration due to
flow leakage of mucus
Filter blocked
Replace filter
Faults 1 - 2 - 3 - 4 - 5 - 6
None of the remedies has
achieved the desired
results
Contact the seller or CA-MI After-sales
Assistance Service
If the overfill security system it’s activated, don’t proceede with the liquid aspiration.
If the overfill security system doesn’t work there are two cases:
1° case – If the overfill security system doesn’t work the aspiration will be stopped by the bacteriological filter
who avoid the liquid penetration inside the device.
2° case – If both the security system doesn’t work, there is the possibility that liquid comes inside the device,
in this case return the device to CA-MI technical service.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING,
PLEASE CONTACT CA-MI TECHNICAL SERVICE.
CA-MI DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL SERVICE
CHECKING, APPEARS TO BE TAMPERED.