![C2AI HD35 Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/c2ai/hd35/hd35_product-information_3130183024.webp)
Product information Wireless data loggers
Technical specifications
HD35AP – HD35APG/HD35AP3G – HD35APS – HD35APW base units
External RF
antenna
LEDs
Power supply
connector
Mini USB
RS485 M12 connector
(HD35APS only)
or
RJ45 Ethernet connector
(HD35APW only)
Versions
HD35AP
: USB output only
HD35APS
: USB and RS485 MODBUS-RTU outputs
HD35APW
: USB output, Wi-Fi and Ethernet interface
HD35APG
:
USB output and GSM/GPRS
module
HD35AP3G
:
USB output and 3G/GSM/GPRS
module
Power supply
Internal 3.7 V lithium-ion
rechargeable
battery,
capacity 2250 mA/h, JST 3-pole connector
Optional
6 Vdc external power adapter (
SWD06
)
Powered directly from the PC USB port
(*)
Power consumption
≈30 mA (E, U) / ≈38 mA (J) without Ethernet/Wi-Fi and
with typical GSM/3G activity
(**)
≈180 mA with Ethernet, ≈150 mA with Wi-Fi
Battery autonomy
(typical)
≈3 days (E, U) / > 2 days (J) if not connected to the
local network and with typical GSM/3G activity
(**)
≈12 hours with Ethernet, ≈14 hours with Wi-Fi
Transmitting frequency
868 MHz, 902-928 MHz or 915.9-929.7 MHz (not for
HD35APG) depending on the model
Antenna
Fixed whip external RF antenna
Whip external GSM/3G antenna with cable (HD35APG
and HD35AP3G only)
Transmitting range
See table 2
Serial outputs
USB with Mini-USB type connector (cable
CP31
)
RS485 with
MODBUS-RTU
protocol (HD35APS only)
Ethernet connection
Only in HD35APW model. Permits (if the Internet
connection is available) sending alarm
and
the recorded data via
, to an
FTP
address or to
an HTTP server (
Cloud
). Allows the
MODBUS TCP/
IP
protocol.
Wi‑Fi connection
Only in HD35APW model. Permits (if the Internet
connection is available) sending alarm
and
the recorded data via
, to an
FTP
address or to
an HTTP server (
Cloud
). Allows the
MODBUS TCP/
IP
protocol.
GSM/GPRS connection
Only in HD35APG and HD35AP3G models. For
sending alarm
or
SMS
and data via
, to
an
FTP
address or to an HTTP server (
Cloud
). Allows
the
GPRS TCP/IP
protocol.
3G connection
Only in HD35AP3G model. For sending alarm
or
SMS
and data via
, to an
FTP
address or
to an HTTP server (
Cloud
). Allows the
3G TCP/IP
protocol.
Internal memory
The number of samples that can be stored depends
on the type of data loggers connected. The capacity
is 226,700 samples if all the data loggers record 7
quantities.
LED indicators
Presence of external power supply, battery charge
level, RF communication status.
Working temperature and
humidity range
-10…+60 °C / 0…85
%RH not condensing
Housing
Plastic material
Dimensions 135 x 86 x 33 mm (excluding antenna)
Weight
200 g approx. (including battery)
Installation
Wall mount support (
supplied
) for removable
installation or flanges (
optional
) for fixed installation
(*)
The connection of the SWD06 external power supply is necessary if the Ethernet, Wi-Fi or GSM/3G transmission is used.
(**)
The intensive use of the GSM/3G transmission can significantly increase the power consumption and reduce the battery life.
24