GasAlertMicroClip
Operator’s Manual

8

Pushbuttons

Pushbutton

Description

C

• To activate the detector, press 

C

.

• To deactivate the detector, press and hold 

C

 until the 

OFF

 countdown is complete and the LCD deactivates

• To view the TWA, STEL, and MAX (maximum) readings, press 

C

 twice. To clear the TWA, STEL, and MAX 

readings, press 

C

 when the LCD displays 

RESET

.

• To initiate calibration, deactivate the detector. Press and hold 

C

 while the detector performs the 

OFF

 

countdown. Continue holding 

C

 while the LCD briefly deactivates and then begins the 

CAL

 countdown. 

Release 

C

 when the 

CAL

 countdown is complete.

• To activate the backlight while in normal operation, press 

C

.

• To acknowledge latched alarms, press 

C

.

• To acknowledge a low alarm and disable the audible alarm, press 

C

 (if the 

Low Alarm Acknowledge

 option 

is enabled).

Содержание Micro Clip XL

Страница 1: ...1 2 3 and 4 Gas Detector Operator Manual...

Страница 2: ...alarm Note The detector is shipped with English as the default displayed language Additional languages provided are Portuguese Spanish German and French The screens for the additional languages are di...

Страница 3: ...in the LEL range calibrate the sensor using the appropriate gas Only the combustible gas detection portion of this instrument has been assessed for performance by CSA International Calibrate only in...

Страница 4: ...e sensor from exposure to lead compounds silicones and chlorinated hydrocarbons Sensor exposure to certain organic vapors such as leaded gasoline and halogenated hydrocarbons may temporarily inhibit s...

Страница 5: ...11 2 2 1 2 Item Description 1 IntelliFlash 2 Visual alarm indicators LEDs 3 Alligator clip 4 Charging connector IR interface 5 Pushbutton 6 Carbon monoxide CO sensor 7 Hydrogen sulfide H2S sensor 8 Ox...

Страница 6: ...lements 7 Display Elements 1 8 7 6 5 4 3 2 Item Description 1 Alarm condition 2 Automatically zero sensor 3 Numeric value 4 Stealth mode 5 Battery life indicator 6 Gas identifier bars 7 Gas cylinder 8...

Страница 7: ...X readings press C when the LCD displays RESET To initiate calibration deactivate the detector Press and hold C while the detector performs the OFF countdown Continue holding C while the LCD briefly d...

Страница 8: ...ol based cleaners Clean the exterior with a soft damp cloth Do not use soaps polishes or solvents Below are common products to avoid using around sensors Cleaners and Lubricants Brake cleaners Lubrica...

Страница 9: ...D then activates again and performs the CAL countdown Continue holding C until the countdown is complete to enter calibration 5 The LCD displays CAL DUE Next a screen displays showing the number of da...

Страница 10: ...GasAlertMicroClip Attach the Gas Cylinder to the Detector 11 Attach the Gas Cylinder to the Detector...

Страница 11: ...nual 2 Connect the calibration hose to the calibration cap 3 Attach the calibration cap to the detector 4 Apply gas Verify the visual and audible alarms activate 5 Close the regulator and remove the c...

Страница 12: ...Vibrator alarm activates TWA Alarm Slow siren Slow alternating flash L and gas bar flash Vibrator alarm activates High Alarm Fast siren Fast alternating flash L and gas bar flash Vibrator alarm activ...

Страница 13: ...activates After 15 minutes of the Low Battery alarm the Automatic Shutdown Alarm sequence begins OFF displays before deactivating Automatic Shutdown Alarm Sequence of 10 rapid sirens and alternating f...

Страница 14: ...Startup When enabled the detector automatically zeros the H2S CO and LEL sensors during the startup self test LEL By Volume CH4 When enabled the LEL reading is displayed as vol assuming a methane env...

Страница 15: ...lliFlash Interval Enter a value 1 60 seconds to define how often IntelliFlash occurs and the detector beeps Intelliflash and or Confidence Beep must be enabled in order to define Confidence Beep and I...

Страница 16: ...of all maintenance bump tests calibrations and alarm events Clean the exterior with a soft damp cloth Do not use solvents soaps or polishes Replacing a Sensor or Sensor Filter a Warning To avoid pers...

Страница 17: ...y Remove the two screws on the PCB Remove the PCB carefully a Caution Ensure no damage occurs to the battery 2 Remove the old sensor filter t may be stuck to the sensors 3 Place the new sensor filter...

Страница 18: ...lace the sensor filter into the front shell 2 Slide the sensors out Note Detectors that are configured for 1 2 or 3 gases may contain a dummy sensor in one of the four sensor locations 3 Insert the ne...

Страница 19: ...plete the detector refer to Reassembling the detector Reassembling the detector Verify the PCB is seated correctly and inserted exactly as it was removed sensors face the front shell Replace the two P...

Страница 20: ...certified by CSA International on the combustible sensor with 5 accuracy Storage temperature 40 C to 50 C 40 F to 122 F Operating humidity 0 to 95 relative humidity non condensing Alarm setpoints May...

Страница 21: ...onfidence beep latching alarms enable disable safe display mode oxygen measurement combustible sensor measurement sensor disable define calibration interval force calibration calibration lock force bu...

Страница 22: ...ertMicroClip charging adapter a Warning Charge only in a safe area that is free of hazardous gas and within temperatures of 0 C to 45 C 32 F to 113 F First time charge XT 2 3 hours XL 5 6 hours Normal...

Страница 23: ...may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to...

Отзывы: