Force Calibration When Overdue

 When enabled, the 

detector automatically enters calibration during startup for 

overdue sensors. If the sensors are not calibrated 

immediately, the detector will deactivate.

Cal Lock

 When enabled, the sensors can only be 

calibrated using an IR device (IR Link with Fleet Manager 

II or the MicroDock II base station).

Force Bump When Overdue

 When enabled, a bump test 

is required if the sensor has exceeded its bump test 

interval. If a successful bump test is not performed, the 

detector will deactivate.

IntelliFlash

 When enabled, the green LED flashes to 

provide continuous visual confirmation that the detector is 

operating correctly. 

IntelliFlash automatically deactivates during a low 
battery alarm, a self-test fail, a calibration fail, a bump 
test fail, or during an alarm event. To define how often 
IntelliFlash occurs (

1

-

60

 seconds), refer to Confidence 

Beep and IntelliFlash Interval. Default setting is 
1 second. Detector is shipped with IntelliFlash enabled.

Confidence Beep and IntelliFlash Interval

 Enter a 

value (

1

-

60

 seconds) to define how often IntelliFlash 

occurs and the detector beeps. Intelliflash and/or 

Confidence Beep must be enabled in order to define 

Confidence Beep and IntelliFlash Interval.

Language

 Select the language to display on the LCD: 

English

Français

 (French), 

Deutsch

 (German), 

Español

 

(Spanish), or 

Português

 (Portuguese).

Charging the Rechargeable Battery

a

 Warning

Only the manufacturer can replace the rechargeable 
battery. Failure to adhere to this caution can lead to fire 
and/or explosion.

Charge only in a safe area that is free of hazardous gas 
and within temperatures of 0°C to 45°C (32°F to 113°F).

The charging adapter is specific to your region. Use of the 
charging adapter outside your region will damage the 
charger and the detector.

Do not calibrate during or immediately after charging.

a

 Avertissement

Seul le fabricant est habilité à remplacer la rechargeable 
batterie. Le non-respect de cette consigne peut provoquer 
un incendie et/ou une explosion.

Chargez le détecteur uniquement dans une zone sûre ne 
renfermant pas de gaz dangereux et dont la température 
ambiante reste comprise entre 0 à 45 °C (32 à 113 °F).

To charge the battery:

1.

Deactivate the detector. Insert the charging adapter 

plug into an AC outlet.

Содержание Micro Clip XL

Страница 1: ...1 2 3 and 4 Gas Detector Operator Manual...

Страница 2: ...alarm Note The detector is shipped with English as the default displayed language Additional languages provided are Portuguese Spanish German and French The screens for the additional languages are di...

Страница 3: ...in the LEL range calibrate the sensor using the appropriate gas Only the combustible gas detection portion of this instrument has been assessed for performance by CSA International Calibrate only in...

Страница 4: ...e sensor from exposure to lead compounds silicones and chlorinated hydrocarbons Sensor exposure to certain organic vapors such as leaded gasoline and halogenated hydrocarbons may temporarily inhibit s...

Страница 5: ...11 2 2 1 2 Item Description 1 IntelliFlash 2 Visual alarm indicators LEDs 3 Alligator clip 4 Charging connector IR interface 5 Pushbutton 6 Carbon monoxide CO sensor 7 Hydrogen sulfide H2S sensor 8 Ox...

Страница 6: ...lements 7 Display Elements 1 8 7 6 5 4 3 2 Item Description 1 Alarm condition 2 Automatically zero sensor 3 Numeric value 4 Stealth mode 5 Battery life indicator 6 Gas identifier bars 7 Gas cylinder 8...

Страница 7: ...X readings press C when the LCD displays RESET To initiate calibration deactivate the detector Press and hold C while the detector performs the OFF countdown Continue holding C while the LCD briefly d...

Страница 8: ...ol based cleaners Clean the exterior with a soft damp cloth Do not use soaps polishes or solvents Below are common products to avoid using around sensors Cleaners and Lubricants Brake cleaners Lubrica...

Страница 9: ...D then activates again and performs the CAL countdown Continue holding C until the countdown is complete to enter calibration 5 The LCD displays CAL DUE Next a screen displays showing the number of da...

Страница 10: ...GasAlertMicroClip Attach the Gas Cylinder to the Detector 11 Attach the Gas Cylinder to the Detector...

Страница 11: ...nual 2 Connect the calibration hose to the calibration cap 3 Attach the calibration cap to the detector 4 Apply gas Verify the visual and audible alarms activate 5 Close the regulator and remove the c...

Страница 12: ...Vibrator alarm activates TWA Alarm Slow siren Slow alternating flash L and gas bar flash Vibrator alarm activates High Alarm Fast siren Fast alternating flash L and gas bar flash Vibrator alarm activ...

Страница 13: ...activates After 15 minutes of the Low Battery alarm the Automatic Shutdown Alarm sequence begins OFF displays before deactivating Automatic Shutdown Alarm Sequence of 10 rapid sirens and alternating f...

Страница 14: ...Startup When enabled the detector automatically zeros the H2S CO and LEL sensors during the startup self test LEL By Volume CH4 When enabled the LEL reading is displayed as vol assuming a methane env...

Страница 15: ...lliFlash Interval Enter a value 1 60 seconds to define how often IntelliFlash occurs and the detector beeps Intelliflash and or Confidence Beep must be enabled in order to define Confidence Beep and I...

Страница 16: ...of all maintenance bump tests calibrations and alarm events Clean the exterior with a soft damp cloth Do not use solvents soaps or polishes Replacing a Sensor or Sensor Filter a Warning To avoid pers...

Страница 17: ...y Remove the two screws on the PCB Remove the PCB carefully a Caution Ensure no damage occurs to the battery 2 Remove the old sensor filter t may be stuck to the sensors 3 Place the new sensor filter...

Страница 18: ...lace the sensor filter into the front shell 2 Slide the sensors out Note Detectors that are configured for 1 2 or 3 gases may contain a dummy sensor in one of the four sensor locations 3 Insert the ne...

Страница 19: ...plete the detector refer to Reassembling the detector Reassembling the detector Verify the PCB is seated correctly and inserted exactly as it was removed sensors face the front shell Replace the two P...

Страница 20: ...certified by CSA International on the combustible sensor with 5 accuracy Storage temperature 40 C to 50 C 40 F to 122 F Operating humidity 0 to 95 relative humidity non condensing Alarm setpoints May...

Страница 21: ...onfidence beep latching alarms enable disable safe display mode oxygen measurement combustible sensor measurement sensor disable define calibration interval force calibration calibration lock force bu...

Страница 22: ...ertMicroClip charging adapter a Warning Charge only in a safe area that is free of hazardous gas and within temperatures of 0 C to 45 C 32 F to 113 F First time charge XT 2 3 hours XL 5 6 hours Normal...

Страница 23: ...may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to...

Отзывы: