High Level Disinfection
It is recommended that you use Cidex
®
OPA by Ethicon US, LLC.
Making sure your probe is compatible with HLD:
1. Tap to display the Settings screen.
2. Tap My iQ to display the My iQ screen.
3. Ensure that the High-Level Disinfection Supported line indicates Yes.
4. Proceed with HLD only if supported on your probe.
5. Disconnect the probe from the mobile device.
Disinfecting the probe using the High-Level Disinfection (HLD) method:
1. After cleaning the probe, you must disinfect the probe. It is recommended that you use Cidex
®
OPA high-level
disinfection solution.
2. Prepare Cidex
®
OPA high-level disinfection solution for use per the manufacturer’s instructions. Fill a tray or ba-
sin with the disinfectant solution at room temperature (minimum temperature of 20°C) to a level allowing immer-
sion of the probe up to the immersion line (the dashed line shown in
Figure 5, “Probe Immersion Line” [36]
.
3. Immerse the probe in Cidex
®
OPA solution up to the immersion line and ensure no air or bubbles are trapped.
Allow soaking according to the manufacturer’s instructions.
4. Thoroughly rinse the probe (up to the immersion line) by immersing it in a large volume of room temperature
critical (purified) water for a minimum of one (1) minute. Remove the probe and discard the rinse water. Do not
reuse the water. Always use fresh volumes of water for each rinse. Repeat this stage two (2) additional times for
a total of three (3) rinses.
5. Thoroughly dry all surfaces of the device using a sterile, lint-free wipe or cloth, changing wipes/cloths when nec-
essary to ensure the device is completely dry. Visually inspect the device to ensure all surfaces are clean and
dry. Repeat the drying steps if any moisture is visible.
6. Once clean and disinfected, visually inspect the probe, strain relief, cable, and connector for signs of damage or
wear.
Figure 5. Probe Immersion Line
Cleaning and Disinfecting the Probe
36
Maintenance