Page 43 of 82
Date: 27.01.06
6.11
Monitor output
The monitor output is primarily used for input configuration using a potentiometer.
This output must not be loaded nor used for control purposes. If used incorrectly,
the Sensor Profibus Module can suffer irreparable damage because this signal is
taken from a point directly before the A/D converter.
6.12
Instrument LED display
(1)
Idle LED (yellow)
•
on/off:
serious error, program execution stopped
•
flashing:
normal program execution
(2)
PROFIBUS LED (green)
•
flashing rapidly:
no baud rate detected
•
flashing slowly:
baud rate detected but no active data transfer
•
on:
device is configured and involved in data transfer
(3)
ERROR LED (red)
•
off:
no error
•
on:
error, e.g. overloaded
•
flashing:
evaluation result frozen
6.13
Test measurement
In the
Test measurement
submenu you can choose to test out the configured
data. The measurement data can be transferred to a PC via the RS 232 interface
and exported to an Excel spreadsheet.