Digital indicator
Model 9163
Page
29
Input IN1 strain gage sensor, 4-wire
Sensor excitation
5 / 10 V
IN1
-
+
-
+
+
-
- Exc
+ Exc
CAL
CAL
Note:
Connect the "CAL“ sensor line to terminal 24 so that it is at the same potential as "-Exc". If the connecting
leads are swapped over, the digital indicator displays the error "
Hi
" or "
E
.
CAL
.x
" after an 80 % calibration.
The test and calibration report for the sensor will specify the size of the calibration resistor.
Input IN2 strain gage sensor, 4-wire
Sensor excitation
5 / 10 V
IN2
-
+
-
+
+
-
- Exc
+ Exc
CAL
CAL
Note:
Connect the "CAL“ sensor line to terminal 26 so that it is at the same potential as "-Exc". If the connecting
leads are swapped over, the digital indicator displays the error "
Hi
" or "
E
.
CAL
.x
" after an 80 % calibration.
The test and calibration report for the sensor will specify the size of the calibration resistor.