10. Disposal
Please fulfill your legal obligations and dispose of unserviceable equipment in
accordance with applicable legal requirements. Thus you contribute to envi-
ronmental protection.
11. Calibration
At the time of delivery, torque sensors have been adjusted and tested with trace-
able calibrated measuring equipment at factory side. Optionally, a calibration
of the sensors can be carried out.
11.1 Proprietary Calibration
Acquisition of measurement points and issuing of a calibration protocol trace-
able calibrated measuring equipment is being used for the calibration. The
sensor data are being checked during this calibration.
11.2 DKD/DAkkS Calibration
The calibration of the sensor is carried out according to the guidelines of the
DKD/DAkkS. The surveillance of the calibrating-laboratory takes place by the
DKD/DAkkS. At this calibration, the uncertainly of measurement of the torque
sensors is determined. Further information can be obtained burster praezi-
sionsmessgeräte gmbh & co kg.
11.3 RE-Calibration
The re-calibration of the torque sensor should be carried out after 26 months.
at the latest.
Shorter intervals are appropriate:
►
overload of the sensor,
►
after repair,
►
after inappropriate handling,
►
demand of high-quality standards,
►
special traceability requirements.