English
500 mPas) from a maximum suction depth of 5 m.
`
`
The instrument is suitable for sampling liquids from
0 - 40 °C.
`
`
For chemically demanding applications use pump
hose made of Marprene (beige, durable), for food
applications use pump hose made of silicone
(transparent).
`
`
The instrument is only suitable to take samples and
not for continuous operation.
`
`
The instrument is suitable for filling open sampling
bottles.
`
`
Only use the instrument with the recommended
sample bottles specified in the accessories.
`
`
Protect the instrument against effects of weather
exposure during storage.
Parts in contact with medium
`
`
PE (Suction hose)
`
`
PP (Hose connector)
`
`
Marprene or silicone (Pump hose)
`
`
LDPE (Sample bottle)
Scope of delivery
`
`
Check delivery for completeness and transport damage.
`
`
Point out any damage in the packaging to the
freight forwarder/supplier immediately.
Assembly
1. Insert the pump hose
into the pump head
.
2. Cut the suction hose
off to the required length and
attach it to the pump hose
with the hose connector
.
3. Fix the bottle basket
in place with the tightener
.
4. Place the sample bottle
in the bottle basket
.
5. Insert the pump hose
about 1 - 3 cm into the
sample bottle
.
6. Place the pump attachment
in the holder
of
the Bosch drive unit
. Turn the locking ring
11
in
the direction
until you hear it snap into place.
Safe sampling
1. Set the gear selector switch
12
to „Gear 1“ (low
speed range).
2. Lower the suction hose
into the sample liquid.
3. Press the On/Off switch
13
to take the sample.
2 x
Pump hose (1x Marprene, 30 cm;
1 x Silicone 30 cm)
1 x
Pump head
1 x
Suction hose , PE, 5 m
1 x
Hose connector, PP
1 x
Bottle basket
1 x
Sample bottle, LDPE, 250 ml
1 x
Bosch drive unit
2 x
Rechargeable battery
14
1 x
Battery charger
15
CAUTION!
The mounting bracket may break.
`
`
Do not load the suspension too heavily (max. 2 kg).
`
`
Check the bottle basket to ensure it is properly
seated. Allow the end of the holding rod to
protrude out about 5 cm upward.
`
`
Wear appropriate protective clothing.
NOTE!
Allow the pump hose to protrude loosely into the
sample bottle.
CAUTION!
Bursting of hoses and hose connectors due to wear
or chemical stress.
`
`
Carry out a visual inspection every time before using.
NOTE!
Follow the safety instructions for the Bosch drive
unit and pump head.
CAUTION!
Possibility of electrical shock.
`
`
Follow the instructions of the drive manufactu-
rer/battery charger manufacturer.
CAUTION!
Risk of atomisation. Ingress of liquids in the instrument.
`
`
In case of uncontrolled escape of hazardous
liquid, interrupt the sampling immediately and
remove the Bosch drive unit.
`
`
Contact the dealer or manufacturer.