61
General settings - Diagnostics
•
To read the events that are related to the product, do the following procedure:
→
Go to device
pH sensor MS01
function
general settings
tab
Diagnostics
.
→
Select
Logbook
.
The events that are related to the product are displayed. Refer to
Symbol is pale: the event
has occurred and is no
more active
pH sensor MS01 | General settings | Diagnostics
19 Messages
20.08.2019 - 13:19:21
20.08.2019 - 13:19:25
Message
büS event: producer not found
Time
15.07.2019 - 17:21:49
Signature
32:45:1793:RMnNsPL:0
20.08.2019 - 11:36:12
20.08.2019 - 11:45:15
15.07.2019 - 17:21:49
29.07.2019 - 07:41:21
29.07.2019 - 07:45:23
29.07.2019 - 07:47:27
Back to previous
view
Event message
Bürkert-internal
code
Symbol is bright: the event
is still active
Event happens during the
start of the product
Time at which the event was checked
Time at which the event occurred
Symbol related to the event
status. Refer to
Fig. 20:
Overview of events in the logbook
Table 15: Description of the symbols
Symbol
Status
Description
Failure, error or
fault
Malfunction
Function check
Ongoing work on the product. For example, simulating measurement
values
Out of
specification
At least one of the monitored parameters is outside its monitored limits.
Maintenance
required
The product is in controlled operation; however, the function is briefly
restricted.
→
Do the required maintenance operation.
Diagnostics
active and no
event has been
generated
Status changes are shown in colour.
Messages are listed and possibly transmitted through any connected
fieldbus.
Diagnostics
inactive
Status changes are not shown.
Messages are neither listed nor transmitted through any connected
fieldbus.
English
Type MS01
Содержание Type MS01
Страница 1: ...Operating Instructions Type MS01 pH sensor cube...
Страница 86: ...86 English Type MS01...
Страница 87: ......
Страница 88: ...www burkert com...