pH sensor - Diagnostics
12.1.3
2-point calibration procedure of the pH sensor
Caution
Risk of injury due to the nature of the operating agents
▶
Respect the prevailing regulations on accident prevention and safety relating to the use of dangerous
fluids.
▶
Observe the information on the safety data sheet. The safety data sheets can be found at:
-
Cleaning solution acid:
-
Cleaning solution alkaline:
-
-
-
-
▶
Wear personal protective equipment when working with cleaning solutions.
▶
Dispose of waste containing chemicals produced by the system in an environmentally friendly manner.
Prerequisites:
•
Buy a Type MZ15 handheld calibration-and-cleaning tool with article number 00568805.
•
Buy a büS extension-cable. Refer to Table 7
. If needed, buy a Y junction.
Table 7: Accessories: cables and connectors
Accessories
Article number
büS extension-cable with 5-pin M12 connectors, 0.5 m
772403
büS extension-cable with 5-pin M12 connectors, 1 m
772404
büS extension-cable with 5-pin M12 connectors, 3 m
772405
büS extension-cable with 5-pin M12 connectors, 5 m
772406
Y junction
772420
•
Buy a buffer solution with pH 5, article number 806698 and a buffer solution with pH 9, article number
806700.
→
Read the Type MZ15 Operating Instructions.
→
Observe the information on the safety data sheets for the used solutions. The relevant data sheet can be
found at the address given on the bottom of the bottle.
→
For a correct use of the Type MZ15 handheld calibration-and-cleaning tool, refer to the Operating
Instructions of the Type MZ15.
→
Unlock and remove the product from the on the backplane of the system.
→
Plug the product on the Type MZ15 handheld calibration-and-cleaning tool. Refer to the Operating
Instructions of the Type MZ15