15
LCD DISPLAY
ACTION
DESCRIPTION
The following functions are for diagnostic and troubleshooting purposes and typically do not have to be adjusted.
Contact your local service agent for assistance before making changes to these functions.
PROGRAMMING (Continued)
CAL LEFT PUMP ?
LEFT
CAL LF PUMP VOL
(-) 58mL (+)
CAL MIDDLE PUMP ?
MIDDLE
CAL MID PUMP VOL
(-) 58mL (+)
CAL RIGHT PUMP ?
RIGHT
CAL RT PUMP VOL
(-) 58mL (+)
CAL HI WTR FLOW
PULL DISPENSER
CAL HI WTR VOL
(-) 1720mL (+)
XXX CAL -> XXX
(-) Tank Temp (+)
BIB EMPTY -> 600
(-) EXIT (+)
Place a 50 mL (#34843.1000)
graduated cylinder under the
dispense tip, then activate the
Left Pump
Use the (+) or (-) buttons to
enter volume collected
Place a 50 mL (#34843.1000)
graduated cylinder under the
dispense tip, then activate the
Middle Pump
Use the (+) or (-) buttons to
enter volume collected
Place a 50 mL (#34843.1000)
graduated cylinder under the
dispense tip, then activate the
Right Pump
Use the (+) or (-) buttons to
enter volume collected
Place a 64oz (2000mL)
container under the dispense
tip, then Press and Hold any
Dispense Switch
Use the (+) or (-) buttons to
enter volume collected
Use the (+) or (-) buttons to
adjust temperature reading
Use the (+) or (-) buttons to
adjust Empty BIB threshold
The Left Pump will dispense concentrate for 10
seconds, then shut off automatically. Collect and
measure the concentrate dispensed.
Enter the volume of concentrate measured from
the Left Pump in mL, (not Oz.)
The Middle Pump will dispense concentrate for
20 seconds, then shut off automatically. Collect
and measure the concentrate dispensed.
Enter the volume of concentrate measured from
the Middle Pump in mL, (not Oz.)
The Right Pump will dispense concentrate for 20
seconds, then shut off automatically. Collect and
measure the concentrate dispensed.
Enter the volume of concentrate measured from
the Right Pump in mL, (not Oz.)
The Dispenser will dispense hot water for 20
seconds, then shut off. Release Dispense Sw.
Collect and measure the water dispensed. Enter
the volume in mL. Default is 1720 mL
Adjust the display reading to match a calibrated
probe inserted next to the Tank Sensor, °F (°C)
Used to set the Conductance Threshold for the
Empty BIB warning. Default is 600
41129 071409