14
MENU SCREEN
ACTION
DESCRIPTION
CAL LF WTR VOL
(-) 36.0OZ (+)
CAL LEFT SIDE ?
EXIT YES
CAL RIGHT SIDE ?
EXIT YES
CAL RIGHT PUMP ?
PULL DISPENSER
CAL RT PUMP VOL
(-) 43mL (+)
CAL RT WTR FLOW
PULL DISPENSER
CAL RT WTR VOL
(-) 36.0OZ (+)
CAL RIGHT SIDE ?
EXIT YES
Use the (+) or (-) buttons to
enter volume collected
NONE
Select (YES) to enable
Calibration routine.
Place a 50 mL (#34843.1000)
graduated cylinder under
the Right dispense tip, then
activate the Right Dispenser
Use the (+) or (-) buttons to
enter volume collected
Place a 64 Oz. (2000 mL)
container under the Right
dispense tip, then activate the
Right Dispenser
Use the (+) or (-) buttons to
enter volume collected
NONE
Enter the volume of the water measured from the
Left dispenser in Oz. (mL).
Scroll down to bypass this screen.
Selecting Yes will enable the Right Dispenser
Calibration routine, (see
Pump Calibration
) prior
to calibrating the dispenser.
The Right Pump will dispense concentrate for
20 seconds, then shut off automatically. Collect
and measure the concentrate dispensed.
Enter the volume of concentrate measured from
the Right Pump in mL(not Oz.) regardless of
Preferred Units selected.
The Right Dispenser will dispense hot water for
20 seconds, then shut off automatically. Collect
and measure the water dispensed.
Enter the volume of the water measured from the
Right dispenser in Oz. (mL)
Scroll down to bypass this screen.
38059 050605