TC-Standard (+) X2
9.5.2 Heat exchanger overview
TC-Standard
Heat exchanger
PTS
PTS-I
2)
PTG
PTG
PTV
PTV-I
2)
MTS
3)
MTS-I
2) 3)
MTG
3)
MTG
3)
MTV
3)
MTV-I
2) 3)
Version / Material
Stainless steel
Glass
PVDF
Stainless steel
Glass
PVDF
Flow rate v
max
1)
450 Nl/h
250 Nl/h
250 Nl/h
300 Nl/h
210 Nl/h
190 Nl/h
Inlet dew point
τ
e,max
1)
65 °C
65 °C
65 °C
65 °C
65 °C
65 °C
Gas inlet temperature
ϑ
G,max
1)
180 °C
140 °C
140 °C
140 °C
140 °C
140 °C
Max. Cooling capacity Q
max
150 kJ/h
90 kJ/h
90 kJ/h
95 kJ/h
80 kJ/h
65 kJ/h
Gas pressure p
max
160 bar
3 bar
2 bar
25 bar
3 bar
2 bar
Pressure drop Δp (v=150 L/h)
10 mbar
10 mbar
10 mbar
20 mbar
19 mbar
18 mbar
Dead volume V
tot
29 ml
29 ml
57 ml
19 ml
18 ml
17 ml
Gas connections (metric)
6 mm
GL 14 (6 mm)
4)
DN 4/6
6 mm tube
GL14 (6 mm)
DN 4/6
Gas connections (US)
1/4"
GL 14 (1/4")
4)
1/4"-1/6"
1/4" tube
GL14 (1/4")
1/4"-1/6"
Condensate out connections (metric)
G3/8
GL 25 (12 mm)
4)
G3/8
G1/4
GL18 (8 mm)
G1/4
Condensate out connections (US)
NPT 3/8"
GL 25 (1/2")
4)
NPT 3/8"
NPT 1/4"
GL18 (8 mm)
NPT 1/4"
1)
Max. cooling capacity of the cooler must be considered
2)
Models marked I have NPT threads or US tubes, respectively.
3)
Passive discharge via automatic condensate drains or traps not applicable for MTG heat exchangers. For passive discharge on
the MTS and MTV heat exchangers, use a screw connection with a clearance of at least 7 mm (see accessories).
4)
Gasket inside diameter
TC-S
Heat exchanger
2x MTG-2
3)
2x MTV-2
3)
2x MTV-2-I
2) 3)
Version / Material
Glass
PVDF
Flow rate v
max
1)
210 Nl/h
190 Nl/h
Inlet dew point
τ
e,max
1)
70 °C
70 °C
Gas inlet temperature
ϑ
G,max
1)
140 °C
140 °C
Max. Cooling capacity Q
max
80 kJ/h
65 kJ/h
Gas pressure p
max
3 bar
2 bar
Pressure drop Δp (v=150 L/h)
19 mbar
18 mbar
Dead volume V
tot
38 ml
36 ml
Gas connections (metric)
GL14 (6 mm)
4)
DN 4/6
Gas connections (US)
GL14 (1/4")
4)
1/4"-1/6"
Condensate out connections (metric)
GL18 (8 mm)
4)
G1/4
Condensate out connections (US)
GL18 (8 mm)
4)
NPT 1/4"
1)
Max. cooling capacity of the cooler must be considered.
2)
Models marked I have NPT threads or US tubes, respectively.
3)
Passive discharge via automatic condensate drains or traps not applicable for MTG-2 heat exchangers. For passive discharge on
the MTV-2 heat exchangers, use a screw connection with a clearance of at least 7 mm (see accessories).
4)
Gasket inside diameter.