GAS 222.21 Ex2
DANGER
Adiabatic compression during gas blowback (explosion hazard)!
Adiabatic compression may cause high gas temperatures and must be checked by the
user.
Gas blowback may result in high gas temperatures due to adiabatic compression. This
can cause flammable gases to ignite spontaneously.
a) Blowback of explosive atmosphere / gases is prohibited.
b) Flammable atmosphere / gases (non-explosive) may only be blown back with nitro-
gen (inert gas).
4.9 Electrical Connections
WARNING
Hazardous electrical voltage
The device must be installed by trained staff only.
CAUTION
Wrong mains voltage
Wrong mains voltage may damage the device.
Regard the correct mains voltage as given on the type plate.
CAUTION
Equipment damage
Cables damaged
Do not damage the cable during installation. Install a strain relief for the cable connec-
tion. Secure the cable against twisting and loosening. Please note the temperature res-
istance of the cables (> 100 °C/212 °F).
Only use cables with a temperature resistance of > 100 °C (212 °F) to connect to power. Make sure the connecting cable has suffi-
cient strain relief (match cable diameter to the seal on the cubic plug/cable fitting).
Please note, the heating system briefly has high starting currents (max. 6 A). Use a suitable fuse (8 A). When connecting, also ob-
serve the applicable explosion protection regulations (e.g. IEC/EN 60079-14).
4.9.1 Version Without Terminal Box
The probe includes two cubic plugs per EN 175301-803. The plug is configured so it cannot be connected reversed. For safety reas-
ons this configuration must not be modified.
One plug is used to power the two heating cartridges (power supply (115/230) VAC, 50/60 Hz, see type plate), the other plug is for
the temperature switch (alarm output).
The supply line cross-sections must be suitable for the rated current. Use a maximum line cross-section of 1.5 mm
2
and a cable
diameter of 8-10 mm.
Connect the power supply and the intrinsically-safe temperature switch per the connection diagram.
Connect optional probe accessories directly to the respective power supply.
4.9.2 Version With Terminal Box
This probe version includes a terminal box. All electrical connections are factory connected to the terminals in the terminal box.
Connect the respective power supply for the two heating cartridges, the intrinsically-safe temperature switch and the optional
accessories to the terminals per the connection diagram.
The probe may only be operated using the Ex e cable fittings and the terminal box closed. Never change the terminal assign-
ment.
13
Bühler Technologies GmbH
BE460048 ◦ 08/2019
Содержание GAS 222.21 Ex2
Страница 4: ...GAS 222 21 Ex2 10 Attached Documents 29 ii Bühler Technologies GmbH BE460048 08 2019 ...
Страница 27: ...GAS 222 21 Ex2 9 6 Dimensions 25 Bühler Technologies GmbH BE460048 08 2019 ...
Страница 28: ...GAS 222 21 Ex2 26 Bühler Technologies GmbH BE460048 08 2019 ...
Страница 35: ......
Страница 36: ......
Страница 37: ......
Страница 38: ......
Страница 39: ......
Страница 48: ...DE 461099 05 2016 Page 8 9 ...