GAS 222.21 Ex1

DANGER

Use in explosive areas

Flammable gas or dust atmospheres could ignite or explode.
Avoid the following hazard sources:

Application area!

Never operate the gas probe outside the specifications.
Sampling gases or gas mixtures which are also explosive in
the absence of air is prohibited.

Electrostatic charge (sparking)!

The equipment may only be used where normal operating
conditions do not frequently produce flammable, electro-
static discharge.

Always clean plastic housing parts and decals with a damp
cloth.

Sparking!

Protect the M3 connectors from external blows.

Flame propagation!

If the process holds a risk of flame propagation, install a
flame arrestor.

Adiabatic compression (explosion hazard)

Adiabatic compression may cause high temperatures during
blowback. Never blowback if explosive gas or dust atmo-
spheres are present. Only use nitrogen (inert gas) to blowback
flammable gas.

Dust

If possible, take the electrical components which must be
opened for repair to a dust-free room. If unable to do so, pre-
vent dust from entering the housing.

Zonal carryover when sampling from zone 20, 21, 22

If the particle size of the dusts being filter is smaller than the
fineness of the filter elements used, zonal carryover from the
process to the probe must be expected.

The fineness of the filter elements used must be clearly smal-
ler than the average particle size of dusts in the process gas.

Dust ignition

Regularly clean dust from all components. Dust layers > 5 mm
may have lower smouldering temperatures and the dust
layer could potentially ignite at a temperature below its
smouldering temperature.

Also remove dust under the thermal insulation and from the
gas probe heating tape.

The ignition temperature and smouldering temperature of
flammable dusts or dust layers which are present must be
above the marked surface temperature plus a safety factor
(also see section "Dust atmosphere requirements").

2.3 Equipment Ambient Temperatures

The ambient temperature range may be limited based on the
version. Please note the Ambient temperature ranges under
"Technical Data” in the full operating instructions.

2.4 Permissible gas inlet temperature

The permissible gas inlet temperatures between versions will
vary based on the temperature class of the gas in den outer
zones (see chapter "Technical Data" in the full operating in-
structions).

3 Transport and storage

Only transport the product inside the original packaging or a
suitable alternative.

The equipment must be protected from moisture and heat
when not in use. They must be stored in a covered, dry and
dust-free room at a temperature between -20 °C to 50 °C (-4 °F
to 122 °F).

4 Installation and connection

NOTICE

Accessories may limit critical operating parameters of
the base unit

Adding accessories may limit critical operating parameters.
Ambient temperatures, zone classifications, explosion
groups, temperature classes or chemical resistances of ac-
cessories may vary from the base unit.

Always include all technical data in the operating instructions
and data sheets of all components in the safety assessment.

4.1 Installation site requirements

Sample gas probes are intended for flange mounting.

• Installation site and installation position are determined

based on requirements specific to the application.

• If necessary, the connection piece should be slightly tilted

toward the centre of the channel.

• The installation site should be protected from the

weather.

• In addition, adequate and safe access for installation and

future maintenance work should be provided. Particularly
follow the uninstalled size of the probe tube!

If the probe is transported to the installation site in pieces, it
will first need to be assembled.

4.2 Installation

DANGER

Danger to life and explosion during installation and
maintenance

The unit must not be worked on (assembly, installation,
maintenance) in explosive atmospheres.

DANGER

Explosion hazard

When used in explosive areas

Flammable gasses and dust could ignite or explode.

Never operate the gas probe outside the specifications. Ex-
tracting gases or gas mixtures which are also explosive in the
absence of air is prohibited.

14

Bühler Technologies GmbH

BX460047 ◦ 03/2021

Содержание GAS 222.21 Ex1

Страница 1: ...60047 Version 03 2021 Ref 998BX460047 Analysentechnik Kurzanleitung Sonden deutsch 2 Brief Instructions Probes english 11 Notice de montage Sondes fran ais 18 Gu a r pida Sondas espa ol 26 chinese sim...

Страница 2: ...ner m glichen Explo sionsgefahr nicht erlaubt Prozessgase oder Prozessgasgemi sche d rfen keine Feststoffe enthalten die in Kombination mit den Materialien der Sonden inklusive Filter und Entnah meroh...

Страница 3: ...tungsbereich des Betreibers unter Zuhilfe nahme seiner Risikobetrachtung Einschr nkungen der Kategorien Zonen durch Zubeh r Es ist zu beachten dass je nach verwendetem Zubeh r der zu gelassene Einsatz...

Страница 4: ...Haftung durch die B hler Techno logies GmbH ausgeschlossen die im Datenblatt und der Anleitung angegebenen Grenz werte eingehalten werden das Griffst ck samt O Ring mit geeignetem Umgebung stemperatur...

Страница 5: ...Beachtung der Vorschrift IEC EN 60079 14 Zus tzliche nationale Bestimmungen bez glich Inbetrieb nahme Betrieb Wartung Instandhaltung und Entsorgung sind einzuhalten GEFAHR Elektrische Spannung Gefahr...

Страница 6: ...izband der Gasentnahmesonde Z ndtemperatur und Glimmtemperatur vorhandener brenn barer St ube bzw Staubschichten m ssen oberhalb der ge kennzeichneten Oberfl chentemperatur liegen und um einen Sicherh...

Страница 7: ...ch am Filterge h use arretiert 4 5 Montage des Eintrittsfilters optional Der Eintrittsfilter falls erforderlich mit der passenden Verl n gerung muss eingeschraubt werden Danach wird die Sonde unter Ve...

Страница 8: ...den Druckluftbeh lter 10 bar 145 psi Der Betriebsdruck reduziert sich je nach Betriebsspannung siehe Typenschild Magnetventil GEFAHR Adiabatische Kompression beim R cksp len von Gas Explosionsgefahr D...

Страница 9: ...alausgleich Statische Aufladung Statische Aufladungen k nnen zu z ndgef hrlichen Funken bildungen f hren Vermeiden Sie statische Aufladung Alle leitf higen Teile der Sonde m ssen geerdet sein Am Geh u...

Страница 10: ...sf hrliche Beschreibung des Ger tes mit Hinweisen zur Fehlersuche und Reparatur finden Sie in der Originalbe triebsanleitung auf der beigef gten CD oder im Internet un ter www buehler technologies com...

Страница 11: ...em perature class for zone 0 1 for zone 1 ATEX II 1G 2G Ex db1 eb mb2 IIC T5 T6 T1 T2 Ga Gb ATEX II 2G Ex db1 eb mb2 IIC T6 T2 Gb IECEx Ex db1 eb mb2 IIC T5 T6 T1 T2 Ga Gb IECEx Ex db1 eb mb2 IIC T6 T...

Страница 12: ...222018 Zone 0 Zone 1 No zone Zone 21 Zone 0 Zone 21 11 Ex1 20 Ex1 21 Ex1 Ceramic sampling tubes item no 4622200205 4622200210 4622200215 Zone 2 Zone 1 No zone Zone 21 Zone 2 Zone 21 Accessories not su...

Страница 13: ...es and operating instructions are available and observed The respective national accident prevention regulations are observed The permissible data and operational conditions are maintained Safety guar...

Страница 14: ...may be limited based on the version Please note the Ambient temperature ranges under Technical Data in the full operating instructions 2 4 Permissible gas inlet temperature The permissible gas inlet t...

Страница 15: ...e application requirements and be weatherproof 4 7 Connecting the Gas Line The sample gas line must be carefully and properly connected using a suitable fitting This table provides an overview of the...

Страница 16: ...s voltage Wrong mains voltage may damage the device Regard the correct mains voltage as given on the type plate CAUTION Equipment damage Cables damaged Do not damage the cable during installation Inst...

Страница 17: ...specifications NOTICE The weather hood must be closed during operation WARNING Housing or component damage Never exceed the maximum working pressure and temperat ure range of the drive DANGER Explosi...

Страница 18: ...sonde de pr l vement de gaz n est pas autoris en raison d un risque d explosion Les gaz ou m langes de gaz in dustriels ne doivent pas contenir de mati res solides qui com bin es aux mat riaux des son...

Страница 19: ...de l exploitant sur la base de son va luation des risques Limitations des cat gories zones du fait des accessoires Sachez que en fonction des accessoires utilis s le domaine d application autoris des...

Страница 20: ...ambiante appropri e et filtre le cas ch ant les dispositifs de surveillance le dispositif de protection sont correctement raccord s les travaux de maintenance et de r paration non d crits dans ce mod...

Страница 21: ...montage installation et main tenance ne doivent tre r alis s qu en absence d atmosph re explosive DANGER Utilisation dans des zones risque d explosion Les atmosph res charg es de gaz ou de poussi res...

Страница 22: ...nant le lieu d installation Les sondes de pr l vement de gaz sont con ues pour un mon tage sur bride Le lieu et la position de montage sont d termin s en fonc tion des conditions d application pertine...

Страница 23: ...ation ATTENTION Risque de rupture Le mat riau isolant peut se briser manipuler avec pr cau tions ne pas le laisser pas tomber Apr s branchement de la conduite de gaz celle ci doit tre bloqu e et fix e...

Страница 24: ...ourvues d un ruban chauffant autor gulant tant branch dans une bo te de connexion selon les d signa tions de bornes dans le bo tier de connexion Afin de garantir une s curit maximale afin entre autres...

Страница 25: ...nt normal n entra nant pas la formation fr quente de d charges lectrostatiques risque d allumage 5 1 Avant la mise en service Avant la mise en service de l appareil v rifiez que qu les raccordements d...

Страница 26: ...gas de muestreo debido al posible riesgo de explosi n Los gases de proceso o las mezclas de gases de proceso no deben contener ning n s lido que en combinaci n con los materiales de las sondas incluid...

Страница 27: ...del operador con ayuda de su evaluaci n de riesgos Limitaciones de categor as zonas mediante accesorios Debe tenerse en cuenta que dependiendo de los accesorios utilizados el rango de aplicaci n permi...

Страница 28: ...La construcci n de instalaciones el ctricas en zonas con riesgo de explosi n requiere el cumplimiento de la norma IEC EN 60079 14 Tambi n deber n respetarse las restantes normativas na cionales relati...

Страница 29: ...ones No se permite la extracci n de gases mezclas de gases o polvos que puedan ser explosivos en presencia de aire Acumulaci n de electricidad est tica formaci n de chispas Los equipos el ctricos solo...

Страница 30: ...por pie zas deber volver a montarse en primer lugar 4 2 Montaje PELIGRO Peligro de muerte y explosi n durante la instalaci n y las tareas de mantenimiento nicamente pueden realizarse trabajos en el d...

Страница 31: ...e an lisis Una vez que todos los conductos est n conectados y se hayan comprobado a prueba de fugas se vuelve a colocar y a fijar el aislante con gran cuidado ADVERTENCIA Salida de gas El gas de medic...

Страница 32: ...da el uso de 30 mA RCD Asimismo se garantizar una protecci n adecuada por medio de un interruptor de protecci n del con ducto en miniatura La funci n de la protecci n el ctrica debe verificarse cada c...

Страница 33: ...cas y la banda calefac tora est n montadas correctamente y sin da os ninguna de las partes de la sonda de gases de muestreo est desmontada los dispositivos de protecci n y control necesarios est n dis...

Страница 34: ...x db1 eb mb2 IIC T5 T6 T1 T2 Ga Gb IECEx Ex db1 eb mb2 IIC T6 T2 Gb 0 21 20 1 ATEX II 1G 2D Ex db1 eb mb2 llC T5 T1 Ga Ex tb mb2 lllC T80 C T226 C Db ATEX II 1D 2G Ex ta lllC T120 C T300 C Da Ex db1 e...

Страница 35: ...0 1 21 0 21 11 Ex1 20 Ex1 21 Ex1 4622200205 4622200210 4622200215 2 1 21 2 21 MIE 3 mJ 2 1 2 T80 C T120 C T130 C T175 C T226 C T300 C 5 mm 155 C 195 C 205 C 250 C 301 C 375 C 120 C 180 C 195 C 263 C 3...

Страница 36: ...GAS 222 21 Ex1 IEC EN 60079 14 a b c d a b c a b c M3 20 21 22 5 mm 2 3 2 4 3 20 C 50 C 4 F bis 122 F 4 36 B hler Technologies GmbH BX460047 03 2021...

Страница 37: ...4 4 O O O 90 4 5 4 6 4 7 GAS 222 PAV01 PAV01 3 2 1 DN65 PN6 DN3 15 0 G3 4 NPT 1 4 G3 8 12 1 6 mm 1 4 NPT 1 4 G1 2 NPT 1 4 G1 8 G1 4 1 1 NPT 1 4 4 7 1 G3 8 4 7 2 6 mm 1 4 6mm 1 4 4 8 37 B hler Technolo...

Страница 38: ...bar 44 psi 10 bar 145 psi a b 4 9 100 C 212 F 100 C 212 F 4 9 1 4 9 2 30 mA RCD 2500 V DC 10 MOhm 4 9 3 IEC 60127 2 1 3 x Ib IEC UN 10 1500 A 4 mm2 IEC EN 60079 14 4 9 4 5 38 B hler Technologies GmbH...

Страница 39: ...GAS 222 21 Ex1 5 1 O 6 CD www buehler technologies com 7 CD www buehler technologies com 39 B hler Technologies GmbH BX460047 03 2021...

Страница 40: ...T6 T2 Gb IECEx Ex db1 eb mb2 IIC T5 T6 T1 T2 Ga Gb IECEx Ex db1 eb mb2 IIC T6 T2 Gb 0 21 20 1 ATEX II 1G 2D Ex db1 eb mb2 llC T5 T1 Ga Ex tb mb2 lllC T80 C T226 C Db ATEX II 1D 2G Ex ta lllC T120 C T...

Страница 41: ...Ex1 30 Ex1 31 Ex1 46222307 46222307F 2 1 20 21 20 1 2 21 11 Ex1 20 Ex1 21 Ex1 46222026 46222026P 2 1 20 21 20 1 2 21 11 Ex1 20 Ex1 21 Ex1 46222001 462220011 46222006 46222004 46222016 46222017 4622201...

Страница 42: ...58 GAS 222 21 Ex1 000092697 462222110149420013111 001 230V valve 24V UC B hler Technologies GmbH Harkortstr 29 D 40880 Ratingen IBExU17ATEX1088X IECEx IBE 17 0024X Ex eb mb IIC T4 Gb II 2G IECEx Ex eb...

Страница 43: ...GAS 222 21 Ex1 IEC EN 60079 14 a b c d a b c a b c M3 43 B hler Technologies GmbH BX460047 03 2021...

Страница 44: ...GAS 222 21 Ex1 20 21 22 5 2 3 2 4 3 20 C 50 C 4 F 122 F 4 4 1 4 2 4 3 44 B hler Technologies GmbH BX460047 03 2021...

Страница 45: ...GAS 222 21 Ex1 4 4 90 4 5 4 6 4 7 GAS 222 PAV01 PAV01 3 2 1 DN65 PN6 DN3 150 G3 4 NPT 1 4 G3 8 12 1 6 1 4 NPT 1 4 G1 2 NPT 1 4 G1 8 G1 4 1 1 NPT 1 4 45 B hler Technologies GmbH BX460047 03 2021...

Страница 46: ...GAS 222 21 Ex1 4 7 1 G3 8 4 7 2 6 1 4 6 1 4 4 8 3 44 psi 10 145 psi a b 4 9 100 C 212 F 100 C 212 F 4 9 1 46 B hler Technologies GmbH BX460047 03 2021...

Страница 47: ...GAS 222 21 Ex1 4 9 2 30 2500 DC 10 M 4 9 3 3 x Ib IEC 60127 2 1 IEC UN 10 1500 4 2 IEC EN 60079 14 4 9 4 5 47 B hler Technologies GmbH BX460047 03 2021...

Страница 48: ...GAS 222 21 Ex1 5 1 6 www buehler technologies com 7 www buehler technologies com 48 B hler Technologies GmbH BX460047 03 2021...

Страница 49: ...ter Blowback solenoid valve Sample gas solenoid valve i i 1 2 Terminal Diagram Terminal Box Limit Switch OPEN CLOSED The connection diagram shows the limit switch box in the intermediate position Swit...

Отзывы: